A study on the use of the Osstell apparatus to evaluate pedicle screw stability

An in-vitro study using micro-CT

Daisuke Nakashima, Ken Ishii, Morio Matsumoto, Masaya Nakamura, Takeo Nagura

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Pull-out force and insertion torque have not been generally used as intraoperative measures for the evaluation of pedicle screw stability because of their invasiveness. On the other hand, resonance frequency analysis is a non-invasive and repeatable technique that has been clinically used in dentistry to evaluate implant stability e.g. by the Osstell apparatus. In this study, the characteristics of the implant stability quotient (ISQ) value obtained by the Osstell apparatus in the field of spinal surgery were investigated. Biomechanical test materials simulating human bone were used to provide a comparative platform for evaluating each fixation strength measure, including pull-out force, insertion torque, and the ISQ value. To perform pull-out force measurement and to repeat pedicle screw insertion and removal, loosening was artificially created, and its effect was investigated. The grade of loosening was quantified on a micro-CT image after pedicle screw removal. In the comparison of the 3 fixation strength measures, the correlations of the ISQ value with the pull-out force (R2 = 0.339 p <0.0001) and the insertion torque (R2 = 0.337 p <0.0001) were lower than the correlation between pull-out force and insertion torque (R2 = 0.918 p <0.0001). On a micro-CT study, the material volume of the internal threads disappeared after destruction of its integrity due to repeated pedicle screw insertion and removal. Material integrity destruction of the internal threads decreased only the pull-out force and the insertion torque, but it did not affect the ISQ value. The ISQ value only decreased when the material volume of the internal threads disappeared, probably because the ISQ value reflects the resistance against a force in the perpendicular direction of the screw, unlike the conventional measures of fixation strength, such as pull-out force and insertion torque, which reflect axial load.

Original languageEnglish
Article numbere0199362.
JournalPLoS One
Volume13
Issue number6
DOIs
Publication statusPublished - 2018 Jun 1

Fingerprint

micro-computed tomography
Torque
screws
in vitro studies
torque
strength (mechanics)
Dentistry
dentistry
In Vitro Techniques
Pedicle Screws
Force measurement
Axial loads
Surgery
Bone and Bones
Bone
surgery
bones

ASJC Scopus subject areas

  • Biochemistry, Genetics and Molecular Biology(all)
  • Agricultural and Biological Sciences(all)

Cite this

A study on the use of the Osstell apparatus to evaluate pedicle screw stability : An in-vitro study using micro-CT. / Nakashima, Daisuke; Ishii, Ken; Matsumoto, Morio; Nakamura, Masaya; Nagura, Takeo.

In: PLoS One, Vol. 13, No. 6, e0199362., 01.06.2018.

Research output: Contribution to journalArticle

@article{8b4548298862475b8ad06dbebda34031,
title = "A study on the use of the Osstell apparatus to evaluate pedicle screw stability: An in-vitro study using micro-CT",
abstract = "Pull-out force and insertion torque have not been generally used as intraoperative measures for the evaluation of pedicle screw stability because of their invasiveness. On the other hand, resonance frequency analysis is a non-invasive and repeatable technique that has been clinically used in dentistry to evaluate implant stability e.g. by the Osstell apparatus. In this study, the characteristics of the implant stability quotient (ISQ) value obtained by the Osstell apparatus in the field of spinal surgery were investigated. Biomechanical test materials simulating human bone were used to provide a comparative platform for evaluating each fixation strength measure, including pull-out force, insertion torque, and the ISQ value. To perform pull-out force measurement and to repeat pedicle screw insertion and removal, loosening was artificially created, and its effect was investigated. The grade of loosening was quantified on a micro-CT image after pedicle screw removal. In the comparison of the 3 fixation strength measures, the correlations of the ISQ value with the pull-out force (R2 = 0.339 p <0.0001) and the insertion torque (R2 = 0.337 p <0.0001) were lower than the correlation between pull-out force and insertion torque (R2 = 0.918 p <0.0001). On a micro-CT study, the material volume of the internal threads disappeared after destruction of its integrity due to repeated pedicle screw insertion and removal. Material integrity destruction of the internal threads decreased only the pull-out force and the insertion torque, but it did not affect the ISQ value. The ISQ value only decreased when the material volume of the internal threads disappeared, probably because the ISQ value reflects the resistance against a force in the perpendicular direction of the screw, unlike the conventional measures of fixation strength, such as pull-out force and insertion torque, which reflect axial load.",
author = "Daisuke Nakashima and Ken Ishii and Morio Matsumoto and Masaya Nakamura and Takeo Nagura",
year = "2018",
month = "6",
day = "1",
doi = "10.1371/journal.pone.0199362",
language = "English",
volume = "13",
journal = "PLoS One",
issn = "1932-6203",
publisher = "Public Library of Science",
number = "6",

}

TY - JOUR

T1 - A study on the use of the Osstell apparatus to evaluate pedicle screw stability

T2 - An in-vitro study using micro-CT

AU - Nakashima, Daisuke

AU - Ishii, Ken

AU - Matsumoto, Morio

AU - Nakamura, Masaya

AU - Nagura, Takeo

PY - 2018/6/1

Y1 - 2018/6/1

N2 - Pull-out force and insertion torque have not been generally used as intraoperative measures for the evaluation of pedicle screw stability because of their invasiveness. On the other hand, resonance frequency analysis is a non-invasive and repeatable technique that has been clinically used in dentistry to evaluate implant stability e.g. by the Osstell apparatus. In this study, the characteristics of the implant stability quotient (ISQ) value obtained by the Osstell apparatus in the field of spinal surgery were investigated. Biomechanical test materials simulating human bone were used to provide a comparative platform for evaluating each fixation strength measure, including pull-out force, insertion torque, and the ISQ value. To perform pull-out force measurement and to repeat pedicle screw insertion and removal, loosening was artificially created, and its effect was investigated. The grade of loosening was quantified on a micro-CT image after pedicle screw removal. In the comparison of the 3 fixation strength measures, the correlations of the ISQ value with the pull-out force (R2 = 0.339 p <0.0001) and the insertion torque (R2 = 0.337 p <0.0001) were lower than the correlation between pull-out force and insertion torque (R2 = 0.918 p <0.0001). On a micro-CT study, the material volume of the internal threads disappeared after destruction of its integrity due to repeated pedicle screw insertion and removal. Material integrity destruction of the internal threads decreased only the pull-out force and the insertion torque, but it did not affect the ISQ value. The ISQ value only decreased when the material volume of the internal threads disappeared, probably because the ISQ value reflects the resistance against a force in the perpendicular direction of the screw, unlike the conventional measures of fixation strength, such as pull-out force and insertion torque, which reflect axial load.

AB - Pull-out force and insertion torque have not been generally used as intraoperative measures for the evaluation of pedicle screw stability because of their invasiveness. On the other hand, resonance frequency analysis is a non-invasive and repeatable technique that has been clinically used in dentistry to evaluate implant stability e.g. by the Osstell apparatus. In this study, the characteristics of the implant stability quotient (ISQ) value obtained by the Osstell apparatus in the field of spinal surgery were investigated. Biomechanical test materials simulating human bone were used to provide a comparative platform for evaluating each fixation strength measure, including pull-out force, insertion torque, and the ISQ value. To perform pull-out force measurement and to repeat pedicle screw insertion and removal, loosening was artificially created, and its effect was investigated. The grade of loosening was quantified on a micro-CT image after pedicle screw removal. In the comparison of the 3 fixation strength measures, the correlations of the ISQ value with the pull-out force (R2 = 0.339 p <0.0001) and the insertion torque (R2 = 0.337 p <0.0001) were lower than the correlation between pull-out force and insertion torque (R2 = 0.918 p <0.0001). On a micro-CT study, the material volume of the internal threads disappeared after destruction of its integrity due to repeated pedicle screw insertion and removal. Material integrity destruction of the internal threads decreased only the pull-out force and the insertion torque, but it did not affect the ISQ value. The ISQ value only decreased when the material volume of the internal threads disappeared, probably because the ISQ value reflects the resistance against a force in the perpendicular direction of the screw, unlike the conventional measures of fixation strength, such as pull-out force and insertion torque, which reflect axial load.

UR - http://www.scopus.com/inward/record.url?scp=85049145467&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85049145467&partnerID=8YFLogxK

U2 - 10.1371/journal.pone.0199362

DO - 10.1371/journal.pone.0199362

M3 - Article

VL - 13

JO - PLoS One

JF - PLoS One

SN - 1932-6203

IS - 6

M1 - e0199362.

ER -