Bipolar radiofrequency ablation: Development of a new expandable device

Nobutake Ito, Jochen Pfeffer, Peter Isfort, Tobias Penzkofer, Christiane K. Kuhl, Andreas H. Mahnken, Thomas Schmitz-Rode, Philipp Bruners

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Purpose: To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. Methods: On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Results: Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 ± 1.5 cm 3 vs. 12.1 ± 1.6 cm3), but were larger than the commercially available bipolar RF probe (vs. 7.3 ± 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 ± 31 s vs. 464 ± 17 s) and shorter than the bipolar probe (vs. 2009 ± 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. Conclusion: The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.

Original languageEnglish
Pages (from-to)770-776
Number of pages7
JournalCardioVascular and Interventional Radiology
Volume37
Issue number3
DOIs
Publication statusPublished - 2014
Externally publishedYes

Fingerprint

Electrodes
Equipment and Supplies
Benchmarking
Liver

Keywords

  • Liver
  • Minimally invasive therapy
  • Radiofrequency ablation

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Cardiology and Cardiovascular Medicine
  • Medicine(all)

Cite this

Ito, N., Pfeffer, J., Isfort, P., Penzkofer, T., Kuhl, C. K., Mahnken, A. H., ... Bruners, P. (2014). Bipolar radiofrequency ablation: Development of a new expandable device. CardioVascular and Interventional Radiology, 37(3), 770-776. https://doi.org/10.1007/s00270-013-0703-2

Bipolar radiofrequency ablation : Development of a new expandable device. / Ito, Nobutake; Pfeffer, Jochen; Isfort, Peter; Penzkofer, Tobias; Kuhl, Christiane K.; Mahnken, Andreas H.; Schmitz-Rode, Thomas; Bruners, Philipp.

In: CardioVascular and Interventional Radiology, Vol. 37, No. 3, 2014, p. 770-776.

Research output: Contribution to journalArticle

Ito, N, Pfeffer, J, Isfort, P, Penzkofer, T, Kuhl, CK, Mahnken, AH, Schmitz-Rode, T & Bruners, P 2014, 'Bipolar radiofrequency ablation: Development of a new expandable device', CardioVascular and Interventional Radiology, vol. 37, no. 3, pp. 770-776. https://doi.org/10.1007/s00270-013-0703-2
Ito, Nobutake ; Pfeffer, Jochen ; Isfort, Peter ; Penzkofer, Tobias ; Kuhl, Christiane K. ; Mahnken, Andreas H. ; Schmitz-Rode, Thomas ; Bruners, Philipp. / Bipolar radiofrequency ablation : Development of a new expandable device. In: CardioVascular and Interventional Radiology. 2014 ; Vol. 37, No. 3. pp. 770-776.
@article{f09be79ec87d4fd0ae2e80c4b48fe0d1,
title = "Bipolar radiofrequency ablation: Development of a new expandable device",
abstract = "Purpose: To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. Methods: On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Results: Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 ± 1.5 cm 3 vs. 12.1 ± 1.6 cm3), but were larger than the commercially available bipolar RF probe (vs. 7.3 ± 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 ± 31 s vs. 464 ± 17 s) and shorter than the bipolar probe (vs. 2009 ± 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. Conclusion: The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.",
keywords = "Liver, Minimally invasive therapy, Radiofrequency ablation",
author = "Nobutake Ito and Jochen Pfeffer and Peter Isfort and Tobias Penzkofer and Kuhl, {Christiane K.} and Mahnken, {Andreas H.} and Thomas Schmitz-Rode and Philipp Bruners",
year = "2014",
doi = "10.1007/s00270-013-0703-2",
language = "English",
volume = "37",
pages = "770--776",
journal = "CardioVascular and Interventional Radiology",
issn = "7415-5101",
publisher = "Springer Verlag",
number = "3",

}

TY - JOUR

T1 - Bipolar radiofrequency ablation

T2 - Development of a new expandable device

AU - Ito, Nobutake

AU - Pfeffer, Jochen

AU - Isfort, Peter

AU - Penzkofer, Tobias

AU - Kuhl, Christiane K.

AU - Mahnken, Andreas H.

AU - Schmitz-Rode, Thomas

AU - Bruners, Philipp

PY - 2014

Y1 - 2014

N2 - Purpose: To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. Methods: On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Results: Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 ± 1.5 cm 3 vs. 12.1 ± 1.6 cm3), but were larger than the commercially available bipolar RF probe (vs. 7.3 ± 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 ± 31 s vs. 464 ± 17 s) and shorter than the bipolar probe (vs. 2009 ± 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. Conclusion: The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.

AB - Purpose: To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. Methods: On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Results: Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 ± 1.5 cm 3 vs. 12.1 ± 1.6 cm3), but were larger than the commercially available bipolar RF probe (vs. 7.3 ± 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 ± 31 s vs. 464 ± 17 s) and shorter than the bipolar probe (vs. 2009 ± 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. Conclusion: The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.

KW - Liver

KW - Minimally invasive therapy

KW - Radiofrequency ablation

UR - http://www.scopus.com/inward/record.url?scp=84901657028&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84901657028&partnerID=8YFLogxK

U2 - 10.1007/s00270-013-0703-2

DO - 10.1007/s00270-013-0703-2

M3 - Article

C2 - 23892758

AN - SCOPUS:84901657028

VL - 37

SP - 770

EP - 776

JO - CardioVascular and Interventional Radiology

JF - CardioVascular and Interventional Radiology

SN - 7415-5101

IS - 3

ER -