Correction of shearing distortions in echo-planar imaging

Guoxiang Liu, Koichi Oshio, Seiji Ogawa, Tsutomu Murata

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

In this study, the existence of linear geometric distortions due to readout (RO) train-induced background gradients was investigated under B0 homogeneity; this paper was aimed at deriving and implementing a simple method for correcting such linear geometric distortions without carrying out the reference measurements of field errors within the echo-planar imaging (EPI) acquisition window. The background gradient in the EPI acquisition window in the RO direction was estimated from the autocorrelation phase, which was calculated from the complex-valued RO 1-D inverse-Fourier-transformed X-κyspace (X-κyspace). All the correction procedures were performed in the complex domain in the hybrid X-κy space after acquiring the data. No prior adjustments or additional measurements are required, and this correction algorithm can be applied to most EPI techniques. The echo-planar images of several phantoms and a human brain were acquired with one 1.5-T and two 3-T MRI systems to validate the effectiveness of the correction method. The severe global spatial shearing distortion caused by the linear background RO gradient in the images was corrected, and improvements were observed in all cases. The results indicate that distortions caused by the background RO gradient under B0 homogeneity can be corrected by online processing, even in high-resolution echo-planar images such as 128 × 128 pixel imaging.

Original languageEnglish
Article number5427123
Pages (from-to)2628-2634
Number of pages7
JournalIEEE Transactions on Magnetics
Volume46
Issue number7
DOIs
Publication statusPublished - 2010 Jul

Fingerprint

Shearing
Imaging techniques
Autocorrelation
Magnetic resonance imaging
Brain
Pixels
Processing

Keywords

  • Distortion
  • Echo-planar imaging (EPI)
  • Eddy currents
  • Magnetic resonance imaging

ASJC Scopus subject areas

  • Electrical and Electronic Engineering
  • Electronic, Optical and Magnetic Materials

Cite this

Correction of shearing distortions in echo-planar imaging. / Liu, Guoxiang; Oshio, Koichi; Ogawa, Seiji; Murata, Tsutomu.

In: IEEE Transactions on Magnetics, Vol. 46, No. 7, 5427123, 07.2010, p. 2628-2634.

Research output: Contribution to journalArticle

Liu, Guoxiang ; Oshio, Koichi ; Ogawa, Seiji ; Murata, Tsutomu. / Correction of shearing distortions in echo-planar imaging. In: IEEE Transactions on Magnetics. 2010 ; Vol. 46, No. 7. pp. 2628-2634.
@article{e143117312ac4fe2922372ef12106ad9,
title = "Correction of shearing distortions in echo-planar imaging",
abstract = "In this study, the existence of linear geometric distortions due to readout (RO) train-induced background gradients was investigated under B0 homogeneity; this paper was aimed at deriving and implementing a simple method for correcting such linear geometric distortions without carrying out the reference measurements of field errors within the echo-planar imaging (EPI) acquisition window. The background gradient in the EPI acquisition window in the RO direction was estimated from the autocorrelation phase, which was calculated from the complex-valued RO 1-D inverse-Fourier-transformed X-κyspace (X-κyspace). All the correction procedures were performed in the complex domain in the hybrid X-κy space after acquiring the data. No prior adjustments or additional measurements are required, and this correction algorithm can be applied to most EPI techniques. The echo-planar images of several phantoms and a human brain were acquired with one 1.5-T and two 3-T MRI systems to validate the effectiveness of the correction method. The severe global spatial shearing distortion caused by the linear background RO gradient in the images was corrected, and improvements were observed in all cases. The results indicate that distortions caused by the background RO gradient under B0 homogeneity can be corrected by online processing, even in high-resolution echo-planar images such as 128 × 128 pixel imaging.",
keywords = "Distortion, Echo-planar imaging (EPI), Eddy currents, Magnetic resonance imaging",
author = "Guoxiang Liu and Koichi Oshio and Seiji Ogawa and Tsutomu Murata",
year = "2010",
month = "7",
doi = "10.1109/TMAG.2010.2041785",
language = "English",
volume = "46",
pages = "2628--2634",
journal = "IEEE Transactions on Magnetics",
issn = "0018-9464",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "7",

}

TY - JOUR

T1 - Correction of shearing distortions in echo-planar imaging

AU - Liu, Guoxiang

AU - Oshio, Koichi

AU - Ogawa, Seiji

AU - Murata, Tsutomu

PY - 2010/7

Y1 - 2010/7

N2 - In this study, the existence of linear geometric distortions due to readout (RO) train-induced background gradients was investigated under B0 homogeneity; this paper was aimed at deriving and implementing a simple method for correcting such linear geometric distortions without carrying out the reference measurements of field errors within the echo-planar imaging (EPI) acquisition window. The background gradient in the EPI acquisition window in the RO direction was estimated from the autocorrelation phase, which was calculated from the complex-valued RO 1-D inverse-Fourier-transformed X-κyspace (X-κyspace). All the correction procedures were performed in the complex domain in the hybrid X-κy space after acquiring the data. No prior adjustments or additional measurements are required, and this correction algorithm can be applied to most EPI techniques. The echo-planar images of several phantoms and a human brain were acquired with one 1.5-T and two 3-T MRI systems to validate the effectiveness of the correction method. The severe global spatial shearing distortion caused by the linear background RO gradient in the images was corrected, and improvements were observed in all cases. The results indicate that distortions caused by the background RO gradient under B0 homogeneity can be corrected by online processing, even in high-resolution echo-planar images such as 128 × 128 pixel imaging.

AB - In this study, the existence of linear geometric distortions due to readout (RO) train-induced background gradients was investigated under B0 homogeneity; this paper was aimed at deriving and implementing a simple method for correcting such linear geometric distortions without carrying out the reference measurements of field errors within the echo-planar imaging (EPI) acquisition window. The background gradient in the EPI acquisition window in the RO direction was estimated from the autocorrelation phase, which was calculated from the complex-valued RO 1-D inverse-Fourier-transformed X-κyspace (X-κyspace). All the correction procedures were performed in the complex domain in the hybrid X-κy space after acquiring the data. No prior adjustments or additional measurements are required, and this correction algorithm can be applied to most EPI techniques. The echo-planar images of several phantoms and a human brain were acquired with one 1.5-T and two 3-T MRI systems to validate the effectiveness of the correction method. The severe global spatial shearing distortion caused by the linear background RO gradient in the images was corrected, and improvements were observed in all cases. The results indicate that distortions caused by the background RO gradient under B0 homogeneity can be corrected by online processing, even in high-resolution echo-planar images such as 128 × 128 pixel imaging.

KW - Distortion

KW - Echo-planar imaging (EPI)

KW - Eddy currents

KW - Magnetic resonance imaging

UR - http://www.scopus.com/inward/record.url?scp=77953841761&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77953841761&partnerID=8YFLogxK

U2 - 10.1109/TMAG.2010.2041785

DO - 10.1109/TMAG.2010.2041785

M3 - Article

AN - SCOPUS:77953841761

VL - 46

SP - 2628

EP - 2634

JO - IEEE Transactions on Magnetics

JF - IEEE Transactions on Magnetics

SN - 0018-9464

IS - 7

M1 - 5427123

ER -