Correlation of effective temporal resolution and accuracy of volume ejection fraction on the ECG-gated cardiac MDCT

Yun Shen, Masahiro Jinzaki, Kozo Sato, Sachio Kuribayashi

Research output: Contribution to journalArticle

Abstract

Purpose: ECG-gated cardiac MDCT offers many cardiac clinical applications. The goals of this report are 1) to introduce a new concept of effective reliability (ER) from effective temporal resolution for the evaluation of ejection fraction (EF) and evaluation of image quality (IQ)of coronary arteries, and 2)to show the correlation of ER and the accuracy of EF with different cardiac reconstruction algorithms and different rotation speeds. Methods and Materials: To assess the accuracy of EF, helical scanning was performed with a gated cardiac MDCT(GE LightSpeed, 8/16 slice)on pulsating cardiac coronary phantoms(0.5 and 0.6 sec rotation speed for each 50-110 bpm, 5-bpm step). We define effective reliability(ER)from effective temporal resolution(%)as follows: ER=(1-TR/HC)x100: TR: Effective temporal resolution, HC: time of heart cycle in each bprn. Results: From the results of the EF measurement and calculated ER, high accuracy was obtained by using optimal scan conditions (optimal rotation speed and cardiac reconstruction algorithm)in a wide range of heart rates (heart rate<90: EF & ER>81%; heart rate<110 bpm: EF & ER>73%). Results showed that the calculated ER is closely correlated with the measurement results of EF based on the phantom experiment (R 2=0.901±0. 075; Max: 0.994, Min: 0.738). Optimal reconstruction thickness can reduce total image number for the evaluation of EF. Conclusion: We concluded that the ER is useful to evaluate EF accuracy and the IQ of images of coronary arteries.

Original languageEnglish
Pages (from-to)233-239
Number of pages7
JournalNippon Acta Radiologica
Volume65
Issue number3
Publication statusPublished - 2005 Jul 25

Fingerprint

Electrocardiography
Coronary Vessels
Heart Rate

Keywords

  • Cardiac phantom
  • Ejection fraction
  • Function
  • MDCT
  • Temporal resolution

ASJC Scopus subject areas

  • Medicine(all)

Cite this

Correlation of effective temporal resolution and accuracy of volume ejection fraction on the ECG-gated cardiac MDCT. / Shen, Yun; Jinzaki, Masahiro; Sato, Kozo; Kuribayashi, Sachio.

In: Nippon Acta Radiologica, Vol. 65, No. 3, 25.07.2005, p. 233-239.

Research output: Contribution to journalArticle

@article{53d0be2e40454374aee96739ac552876,
title = "Correlation of effective temporal resolution and accuracy of volume ejection fraction on the ECG-gated cardiac MDCT",
abstract = "Purpose: ECG-gated cardiac MDCT offers many cardiac clinical applications. The goals of this report are 1) to introduce a new concept of effective reliability (ER) from effective temporal resolution for the evaluation of ejection fraction (EF) and evaluation of image quality (IQ)of coronary arteries, and 2)to show the correlation of ER and the accuracy of EF with different cardiac reconstruction algorithms and different rotation speeds. Methods and Materials: To assess the accuracy of EF, helical scanning was performed with a gated cardiac MDCT(GE LightSpeed, 8/16 slice)on pulsating cardiac coronary phantoms(0.5 and 0.6 sec rotation speed for each 50-110 bpm, 5-bpm step). We define effective reliability(ER)from effective temporal resolution({\%})as follows: ER=(1-TR/HC)x100: TR: Effective temporal resolution, HC: time of heart cycle in each bprn. Results: From the results of the EF measurement and calculated ER, high accuracy was obtained by using optimal scan conditions (optimal rotation speed and cardiac reconstruction algorithm)in a wide range of heart rates (heart rate<90: EF & ER>81{\%}; heart rate<110 bpm: EF & ER>73{\%}). Results showed that the calculated ER is closely correlated with the measurement results of EF based on the phantom experiment (R 2=0.901±0. 075; Max: 0.994, Min: 0.738). Optimal reconstruction thickness can reduce total image number for the evaluation of EF. Conclusion: We concluded that the ER is useful to evaluate EF accuracy and the IQ of images of coronary arteries.",
keywords = "Cardiac phantom, Ejection fraction, Function, MDCT, Temporal resolution",
author = "Yun Shen and Masahiro Jinzaki and Kozo Sato and Sachio Kuribayashi",
year = "2005",
month = "7",
day = "25",
language = "English",
volume = "65",
pages = "233--239",
journal = "Nippon Acta Radiologica",
issn = "0048-0428",
publisher = "Japan Radiological Society",
number = "3",

}

TY - JOUR

T1 - Correlation of effective temporal resolution and accuracy of volume ejection fraction on the ECG-gated cardiac MDCT

AU - Shen, Yun

AU - Jinzaki, Masahiro

AU - Sato, Kozo

AU - Kuribayashi, Sachio

PY - 2005/7/25

Y1 - 2005/7/25

N2 - Purpose: ECG-gated cardiac MDCT offers many cardiac clinical applications. The goals of this report are 1) to introduce a new concept of effective reliability (ER) from effective temporal resolution for the evaluation of ejection fraction (EF) and evaluation of image quality (IQ)of coronary arteries, and 2)to show the correlation of ER and the accuracy of EF with different cardiac reconstruction algorithms and different rotation speeds. Methods and Materials: To assess the accuracy of EF, helical scanning was performed with a gated cardiac MDCT(GE LightSpeed, 8/16 slice)on pulsating cardiac coronary phantoms(0.5 and 0.6 sec rotation speed for each 50-110 bpm, 5-bpm step). We define effective reliability(ER)from effective temporal resolution(%)as follows: ER=(1-TR/HC)x100: TR: Effective temporal resolution, HC: time of heart cycle in each bprn. Results: From the results of the EF measurement and calculated ER, high accuracy was obtained by using optimal scan conditions (optimal rotation speed and cardiac reconstruction algorithm)in a wide range of heart rates (heart rate<90: EF & ER>81%; heart rate<110 bpm: EF & ER>73%). Results showed that the calculated ER is closely correlated with the measurement results of EF based on the phantom experiment (R 2=0.901±0. 075; Max: 0.994, Min: 0.738). Optimal reconstruction thickness can reduce total image number for the evaluation of EF. Conclusion: We concluded that the ER is useful to evaluate EF accuracy and the IQ of images of coronary arteries.

AB - Purpose: ECG-gated cardiac MDCT offers many cardiac clinical applications. The goals of this report are 1) to introduce a new concept of effective reliability (ER) from effective temporal resolution for the evaluation of ejection fraction (EF) and evaluation of image quality (IQ)of coronary arteries, and 2)to show the correlation of ER and the accuracy of EF with different cardiac reconstruction algorithms and different rotation speeds. Methods and Materials: To assess the accuracy of EF, helical scanning was performed with a gated cardiac MDCT(GE LightSpeed, 8/16 slice)on pulsating cardiac coronary phantoms(0.5 and 0.6 sec rotation speed for each 50-110 bpm, 5-bpm step). We define effective reliability(ER)from effective temporal resolution(%)as follows: ER=(1-TR/HC)x100: TR: Effective temporal resolution, HC: time of heart cycle in each bprn. Results: From the results of the EF measurement and calculated ER, high accuracy was obtained by using optimal scan conditions (optimal rotation speed and cardiac reconstruction algorithm)in a wide range of heart rates (heart rate<90: EF & ER>81%; heart rate<110 bpm: EF & ER>73%). Results showed that the calculated ER is closely correlated with the measurement results of EF based on the phantom experiment (R 2=0.901±0. 075; Max: 0.994, Min: 0.738). Optimal reconstruction thickness can reduce total image number for the evaluation of EF. Conclusion: We concluded that the ER is useful to evaluate EF accuracy and the IQ of images of coronary arteries.

KW - Cardiac phantom

KW - Ejection fraction

KW - Function

KW - MDCT

KW - Temporal resolution

UR - http://www.scopus.com/inward/record.url?scp=26944501655&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=26944501655&partnerID=8YFLogxK

M3 - Article

C2 - 16119784

AN - SCOPUS:26944501655

VL - 65

SP - 233

EP - 239

JO - Nippon Acta Radiologica

JF - Nippon Acta Radiologica

SN - 0048-0428

IS - 3

ER -