Abstract
High-performance liquid chromatography with column switching and mass spectrometry (MS) was applied to the on-line determination and resolution of the enantiomers of E2020 (acetylcholinesterase inhibitor) in plasma. This system employs two avidin columns and fast atom bombardment (FAB)-MS. A plasma sample was injected directly into an avidin trapping column (10 mm × 40 mm I.D.). The plasma protein was washed out from the trapping column immediately while E2020 was retained. After the column-switching procedure, E2020 was separated enantioselectivity in an avidin analytical column. The separated E2020 enantiomers were specifically detected by FAB-MS without interference from metabolites of E2020 and plasma constituents. The limit of quantification for each enantiomer of E2020 in plasma was 1.0 ng/ml and the intra- and inter-assay relative standard deviations for the method were less than 5.2%. The assay was validated for enantioselective pharmacokinetic studies in the dog.
Original language | English |
---|---|
Pages (from-to) | 209-218 |
Number of pages | 10 |
Journal | Journal of Chromatography A |
Volume | 694 |
Issue number | 1 |
DOIs | |
Publication status | Published - 1995 Mar 3 |
ASJC Scopus subject areas
- Analytical Chemistry
- Biochemistry
- Organic Chemistry