TY - GEN
T1 - EEG-Hat with Candle-like Microneedle Electrode
AU - Kawana, Takumi
AU - Yoshida, Yuri
AU - Kudo, Yuta
AU - Miki, Norihisa
N1 - Funding Information:
This work was supported by JST PRESTO (Information Environment and Humans), JSPS KAKENHI Grant-in-Aid for Scientific Research (C)17K02347.
PY - 2019/7
Y1 - 2019/7
N2 - This paper reports the EEG-Hat, the hat-type EEG device with candle-like dry microneedle electrodes. Current wearable EEG devices have two major problems; 1) no adjustability to every participant and 2) no ability to measure EEG at hairy part. If these problems are solved, health care technologies in daily life will be developed drastically. The EEG-Hat potentially solves these problems. This device can adjust the electrodes to different size and shape of human head. This device has shutter-like structures to separate hair, so using this device, it is possible to measure EEG without separating hair by hand. First, we quantified the force to press the electrodes and the amount of separating hair. Then, the EEG-Hat fabricated based on the findings successfully measured EEG from 3 hairy parts without manual separation.
AB - This paper reports the EEG-Hat, the hat-type EEG device with candle-like dry microneedle electrodes. Current wearable EEG devices have two major problems; 1) no adjustability to every participant and 2) no ability to measure EEG at hairy part. If these problems are solved, health care technologies in daily life will be developed drastically. The EEG-Hat potentially solves these problems. This device can adjust the electrodes to different size and shape of human head. This device has shutter-like structures to separate hair, so using this device, it is possible to measure EEG without separating hair by hand. First, we quantified the force to press the electrodes and the amount of separating hair. Then, the EEG-Hat fabricated based on the findings successfully measured EEG from 3 hairy parts without manual separation.
UR - http://www.scopus.com/inward/record.url?scp=85077892870&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85077892870&partnerID=8YFLogxK
U2 - 10.1109/EMBC.2019.8857477
DO - 10.1109/EMBC.2019.8857477
M3 - Conference contribution
C2 - 31946088
AN - SCOPUS:85077892870
T3 - Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
SP - 1111
EP - 1114
BT - 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2019
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2019
Y2 - 23 July 2019 through 27 July 2019
ER -