Electrochemical detection of tyrosine derivatives and protein tyrosine kinase activity using boron-doped diamond electrodes

Masanobu Chiku, Kenichi Horisawa, Nobuhide Doi, Hiroshi Yanagawa, Yasuaki Einaga

Research output: Contribution to journalArticle

27 Citations (Scopus)

Abstract

In this report, we determined protein tyrosine kinase (PTKs) activity in human epidermoid carcinoma cells (A431) by employing a novel electrochemical method using boron-doped diamond (BDD) electrodes that enables the electrochemical oxidation of tyrosine (Tyr), phosphorylated Tyr (Tyr-P) and sulfated Tyr (Tyr-S) in water-based solutions. Cyclic voltammetry for Tyr, Tyr-P and Tyr-S showed well-defined oxidation peaks at 0.8. V for Tyr, 1.4. V for Tyr-P and 1.7. V for Tyr-S, respectively. Very little work has been reported previously on the detection of Tyr-P and Tyr-S, probably due to their high oxidation potentials. We utilized electrochemical methods for the detection of kinase activity in connection with poly(Glu-Tyr) modified magnetic beads. Linear-sweep voltammograms for the electrochemical detection of PTKs activity were carried out using BDD electrodes consisting of peptide-modified magnetic beads. Without phosphorylation of the peptide-modified magnetic beads using PTKs, we observed clear oxidation peaks for Tyr oxidation and no significant electrochemical responses for Tyr-P oxidation at 1.4. V for the background. On the other hand, with phosphorylation of the beads using PTKs, the peak oxidation current at 1.4. V clearly increased, while the peak oxidation current for Tyr oxidation decreased. This indicates that PTKs activity could be successfully detected by using electrochemical methods employing BDD electrodes. This method was utilized for the in vitro kinase activity detection of human cell lysate, and the electrochemical measurements were compatible with the Enzyme-Linked ImmunoSorbent Assay based method. Our results indicate that the electrochemical method can be applied to real samples such as cell lysate.

Original languageEnglish
Pages (from-to)235-240
Number of pages6
JournalBiosensors and Bioelectronics
Volume26
Issue number1
DOIs
Publication statusPublished - 2010 Sep

Fingerprint

Diamond
Boron
Protein-Tyrosine Kinases
Tyrosine
Diamonds
Electrodes
Derivatives
Proteins
Oxidation
Phosphorylation
Peptides
Cells
Electrochemical oxidation
Human Activities
Cyclic voltammetry
Phosphotransferases
Assays
Enzymes
Immunosorbents

Keywords

  • Boron-doped diamond electrodes
  • Phosphorylation
  • Protein
  • Protein tyrosine kinase

ASJC Scopus subject areas

  • Biophysics
  • Biomedical Engineering
  • Biotechnology
  • Electrochemistry

Cite this

Electrochemical detection of tyrosine derivatives and protein tyrosine kinase activity using boron-doped diamond electrodes. / Chiku, Masanobu; Horisawa, Kenichi; Doi, Nobuhide; Yanagawa, Hiroshi; Einaga, Yasuaki.

In: Biosensors and Bioelectronics, Vol. 26, No. 1, 09.2010, p. 235-240.

Research output: Contribution to journalArticle

@article{b51d5259169d49ce93a443c456b2d1a9,
title = "Electrochemical detection of tyrosine derivatives and protein tyrosine kinase activity using boron-doped diamond electrodes",
abstract = "In this report, we determined protein tyrosine kinase (PTKs) activity in human epidermoid carcinoma cells (A431) by employing a novel electrochemical method using boron-doped diamond (BDD) electrodes that enables the electrochemical oxidation of tyrosine (Tyr), phosphorylated Tyr (Tyr-P) and sulfated Tyr (Tyr-S) in water-based solutions. Cyclic voltammetry for Tyr, Tyr-P and Tyr-S showed well-defined oxidation peaks at 0.8. V for Tyr, 1.4. V for Tyr-P and 1.7. V for Tyr-S, respectively. Very little work has been reported previously on the detection of Tyr-P and Tyr-S, probably due to their high oxidation potentials. We utilized electrochemical methods for the detection of kinase activity in connection with poly(Glu-Tyr) modified magnetic beads. Linear-sweep voltammograms for the electrochemical detection of PTKs activity were carried out using BDD electrodes consisting of peptide-modified magnetic beads. Without phosphorylation of the peptide-modified magnetic beads using PTKs, we observed clear oxidation peaks for Tyr oxidation and no significant electrochemical responses for Tyr-P oxidation at 1.4. V for the background. On the other hand, with phosphorylation of the beads using PTKs, the peak oxidation current at 1.4. V clearly increased, while the peak oxidation current for Tyr oxidation decreased. This indicates that PTKs activity could be successfully detected by using electrochemical methods employing BDD electrodes. This method was utilized for the in vitro kinase activity detection of human cell lysate, and the electrochemical measurements were compatible with the Enzyme-Linked ImmunoSorbent Assay based method. Our results indicate that the electrochemical method can be applied to real samples such as cell lysate.",
keywords = "Boron-doped diamond electrodes, Phosphorylation, Protein, Protein tyrosine kinase",
author = "Masanobu Chiku and Kenichi Horisawa and Nobuhide Doi and Hiroshi Yanagawa and Yasuaki Einaga",
year = "2010",
month = "9",
doi = "10.1016/j.bios.2010.06.027",
language = "English",
volume = "26",
pages = "235--240",
journal = "Biosensors",
issn = "0956-5663",
publisher = "Elsevier Limited",
number = "1",

}

TY - JOUR

T1 - Electrochemical detection of tyrosine derivatives and protein tyrosine kinase activity using boron-doped diamond electrodes

AU - Chiku, Masanobu

AU - Horisawa, Kenichi

AU - Doi, Nobuhide

AU - Yanagawa, Hiroshi

AU - Einaga, Yasuaki

PY - 2010/9

Y1 - 2010/9

N2 - In this report, we determined protein tyrosine kinase (PTKs) activity in human epidermoid carcinoma cells (A431) by employing a novel electrochemical method using boron-doped diamond (BDD) electrodes that enables the electrochemical oxidation of tyrosine (Tyr), phosphorylated Tyr (Tyr-P) and sulfated Tyr (Tyr-S) in water-based solutions. Cyclic voltammetry for Tyr, Tyr-P and Tyr-S showed well-defined oxidation peaks at 0.8. V for Tyr, 1.4. V for Tyr-P and 1.7. V for Tyr-S, respectively. Very little work has been reported previously on the detection of Tyr-P and Tyr-S, probably due to their high oxidation potentials. We utilized electrochemical methods for the detection of kinase activity in connection with poly(Glu-Tyr) modified magnetic beads. Linear-sweep voltammograms for the electrochemical detection of PTKs activity were carried out using BDD electrodes consisting of peptide-modified magnetic beads. Without phosphorylation of the peptide-modified magnetic beads using PTKs, we observed clear oxidation peaks for Tyr oxidation and no significant electrochemical responses for Tyr-P oxidation at 1.4. V for the background. On the other hand, with phosphorylation of the beads using PTKs, the peak oxidation current at 1.4. V clearly increased, while the peak oxidation current for Tyr oxidation decreased. This indicates that PTKs activity could be successfully detected by using electrochemical methods employing BDD electrodes. This method was utilized for the in vitro kinase activity detection of human cell lysate, and the electrochemical measurements were compatible with the Enzyme-Linked ImmunoSorbent Assay based method. Our results indicate that the electrochemical method can be applied to real samples such as cell lysate.

AB - In this report, we determined protein tyrosine kinase (PTKs) activity in human epidermoid carcinoma cells (A431) by employing a novel electrochemical method using boron-doped diamond (BDD) electrodes that enables the electrochemical oxidation of tyrosine (Tyr), phosphorylated Tyr (Tyr-P) and sulfated Tyr (Tyr-S) in water-based solutions. Cyclic voltammetry for Tyr, Tyr-P and Tyr-S showed well-defined oxidation peaks at 0.8. V for Tyr, 1.4. V for Tyr-P and 1.7. V for Tyr-S, respectively. Very little work has been reported previously on the detection of Tyr-P and Tyr-S, probably due to their high oxidation potentials. We utilized electrochemical methods for the detection of kinase activity in connection with poly(Glu-Tyr) modified magnetic beads. Linear-sweep voltammograms for the electrochemical detection of PTKs activity were carried out using BDD electrodes consisting of peptide-modified magnetic beads. Without phosphorylation of the peptide-modified magnetic beads using PTKs, we observed clear oxidation peaks for Tyr oxidation and no significant electrochemical responses for Tyr-P oxidation at 1.4. V for the background. On the other hand, with phosphorylation of the beads using PTKs, the peak oxidation current at 1.4. V clearly increased, while the peak oxidation current for Tyr oxidation decreased. This indicates that PTKs activity could be successfully detected by using electrochemical methods employing BDD electrodes. This method was utilized for the in vitro kinase activity detection of human cell lysate, and the electrochemical measurements were compatible with the Enzyme-Linked ImmunoSorbent Assay based method. Our results indicate that the electrochemical method can be applied to real samples such as cell lysate.

KW - Boron-doped diamond electrodes

KW - Phosphorylation

KW - Protein

KW - Protein tyrosine kinase

UR - http://www.scopus.com/inward/record.url?scp=80052538001&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=80052538001&partnerID=8YFLogxK

U2 - 10.1016/j.bios.2010.06.027

DO - 10.1016/j.bios.2010.06.027

M3 - Article

C2 - 20634053

AN - SCOPUS:80052538001

VL - 26

SP - 235

EP - 240

JO - Biosensors

JF - Biosensors

SN - 0956-5663

IS - 1

ER -