Influence of backscatter radiation on cranial bone fixation devices

Research output: Contribution to journalArticle

Abstract

Postoperative radiation can cause ulcer formation, leading to the denudation of skin over alloplastic materials. The influence of backscatter radiation from fixation devices has not been investigated. The aim of this study is to evaluate backscatter dose variations for different cranial bone fixation devices in an experimental model designed to simulate postoperative radiotherapy. The authors assessed the radiation backscatter doses associated with resorbable (PLLA-PGA) and titanium plates. The samples were irradiated with 6 and 10 MV photon beams from a linear accelerator. Measurements were obtained using an ionization chamber and radiochromic films cut from the same batch. As a result, the backscatter radiation of water and PLLA-PGA proportionally decreased as the depth increased. However, the backscatter radiation of the titanium plate increased just above the plate. This depth lies in the region of the scalp. Each material showed a dose of radioactivity that was higher at 10 MV than that at 6 MV. These devices showed a significant difference, which suggested that these materials amplified the dose compared with water at 6 MV. In conclusion, it is supposed that PLLA-PGA should be used to fix the cranium to decrease the potential for radiation ulcers.

Original languageEnglish
Pages (from-to)1094-1096
Number of pages3
JournalJournal of Craniofacial Surgery
Volume29
Issue number4
DOIs
Publication statusPublished - 2018 Jun 1

Fingerprint

Radiation
Prostaglandins A
Bone and Bones
Equipment and Supplies
Titanium
Ulcer
Particle Accelerators
Water
Scalp
Photons
Skull
Radioactivity
Theoretical Models
Radiotherapy
Skin

Keywords

  • Backscatter
  • PLLA
  • resorbable plate
  • skin ulcer
  • titanium

ASJC Scopus subject areas

  • Surgery
  • Otorhinolaryngology

Cite this

Influence of backscatter radiation on cranial bone fixation devices. / Sakamoto, Yoshiaki; Koike, Naoyoshi; Takei, Hideyuki; Ohno, Mari; Shigematsu, Naoyuki; Kishi, Kazuo.

In: Journal of Craniofacial Surgery, Vol. 29, No. 4, 01.06.2018, p. 1094-1096.

Research output: Contribution to journalArticle

@article{9198707bfeee4237a66605769fa177e3,
title = "Influence of backscatter radiation on cranial bone fixation devices",
abstract = "Postoperative radiation can cause ulcer formation, leading to the denudation of skin over alloplastic materials. The influence of backscatter radiation from fixation devices has not been investigated. The aim of this study is to evaluate backscatter dose variations for different cranial bone fixation devices in an experimental model designed to simulate postoperative radiotherapy. The authors assessed the radiation backscatter doses associated with resorbable (PLLA-PGA) and titanium plates. The samples were irradiated with 6 and 10 MV photon beams from a linear accelerator. Measurements were obtained using an ionization chamber and radiochromic films cut from the same batch. As a result, the backscatter radiation of water and PLLA-PGA proportionally decreased as the depth increased. However, the backscatter radiation of the titanium plate increased just above the plate. This depth lies in the region of the scalp. Each material showed a dose of radioactivity that was higher at 10 MV than that at 6 MV. These devices showed a significant difference, which suggested that these materials amplified the dose compared with water at 6 MV. In conclusion, it is supposed that PLLA-PGA should be used to fix the cranium to decrease the potential for radiation ulcers.",
keywords = "Backscatter, PLLA, resorbable plate, skin ulcer, titanium",
author = "Yoshiaki Sakamoto and Naoyoshi Koike and Hideyuki Takei and Mari Ohno and Naoyuki Shigematsu and Kazuo Kishi",
year = "2018",
month = "6",
day = "1",
doi = "10.1097/SCS.0000000000004392",
language = "English",
volume = "29",
pages = "1094--1096",
journal = "Journal of Craniofacial Surgery",
issn = "1049-2275",
publisher = "Lippincott Williams and Wilkins",
number = "4",

}

TY - JOUR

T1 - Influence of backscatter radiation on cranial bone fixation devices

AU - Sakamoto, Yoshiaki

AU - Koike, Naoyoshi

AU - Takei, Hideyuki

AU - Ohno, Mari

AU - Shigematsu, Naoyuki

AU - Kishi, Kazuo

PY - 2018/6/1

Y1 - 2018/6/1

N2 - Postoperative radiation can cause ulcer formation, leading to the denudation of skin over alloplastic materials. The influence of backscatter radiation from fixation devices has not been investigated. The aim of this study is to evaluate backscatter dose variations for different cranial bone fixation devices in an experimental model designed to simulate postoperative radiotherapy. The authors assessed the radiation backscatter doses associated with resorbable (PLLA-PGA) and titanium plates. The samples were irradiated with 6 and 10 MV photon beams from a linear accelerator. Measurements were obtained using an ionization chamber and radiochromic films cut from the same batch. As a result, the backscatter radiation of water and PLLA-PGA proportionally decreased as the depth increased. However, the backscatter radiation of the titanium plate increased just above the plate. This depth lies in the region of the scalp. Each material showed a dose of radioactivity that was higher at 10 MV than that at 6 MV. These devices showed a significant difference, which suggested that these materials amplified the dose compared with water at 6 MV. In conclusion, it is supposed that PLLA-PGA should be used to fix the cranium to decrease the potential for radiation ulcers.

AB - Postoperative radiation can cause ulcer formation, leading to the denudation of skin over alloplastic materials. The influence of backscatter radiation from fixation devices has not been investigated. The aim of this study is to evaluate backscatter dose variations for different cranial bone fixation devices in an experimental model designed to simulate postoperative radiotherapy. The authors assessed the radiation backscatter doses associated with resorbable (PLLA-PGA) and titanium plates. The samples were irradiated with 6 and 10 MV photon beams from a linear accelerator. Measurements were obtained using an ionization chamber and radiochromic films cut from the same batch. As a result, the backscatter radiation of water and PLLA-PGA proportionally decreased as the depth increased. However, the backscatter radiation of the titanium plate increased just above the plate. This depth lies in the region of the scalp. Each material showed a dose of radioactivity that was higher at 10 MV than that at 6 MV. These devices showed a significant difference, which suggested that these materials amplified the dose compared with water at 6 MV. In conclusion, it is supposed that PLLA-PGA should be used to fix the cranium to decrease the potential for radiation ulcers.

KW - Backscatter

KW - PLLA

KW - resorbable plate

KW - skin ulcer

KW - titanium

UR - http://www.scopus.com/inward/record.url?scp=85048101097&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85048101097&partnerID=8YFLogxK

U2 - 10.1097/SCS.0000000000004392

DO - 10.1097/SCS.0000000000004392

M3 - Article

C2 - 29498970

AN - SCOPUS:85048101097

VL - 29

SP - 1094

EP - 1096

JO - Journal of Craniofacial Surgery

JF - Journal of Craniofacial Surgery

SN - 1049-2275

IS - 4

ER -