Measurement of depth dose distribution using plastic scintillator

Masatoshi Hashimoto, Kiyoyuki Kodama, Takashi Hanada, Tatsuya Ide, Tomoko Tsukahara, Mitsuo Ishikawa, Koichi Maruyama

Research output: Contribution to journalArticle

Abstract

We propose a new method to measure the depth dose distribution in matter. The method is to use organic plastic scintillator as a phantom, which has the effective atomic number close to that of human soft tissue, and to measure the position distribution of the scintillation light by a wavelength analyzer through a thread of plastic optical fiber. The purpose of this study is to examine possibility to measure the depth dose distribution with the proposed method. We carried out the following three measurements. First, we measured the effective field of view (FOV); the plastic optical fiber accepts the light from the FOV. Then, we measured the dose dependence of the amount of emission light. Finally, the light distribution in the depth direction by the present method was compared to the depth dose distribution by the ionization chamber. The results are as follows. The angular range of FOV is 7.03 degrees +/- 0.21 degrees The dose dependence shows a good linearity. In other words, the amount of emission light is proportional to the absorbed dose. Both of the light and the dose distributions are proportional each other, and they show the same attenuation in the region deeper than 15 mm. From these results, we conclude that the present method can be used for the measurement of the depth dose distribution in matter.

Original languageEnglish
Pages (from-to)1424-1431
Number of pages8
JournalNippon Hoshasen Gijutsu Gakkai zasshi
Volume59
Issue number11
Publication statusPublished - 2003
Externally publishedYes

Fingerprint

Plastics
Light
Optical Fibers

Cite this

Hashimoto, M., Kodama, K., Hanada, T., Ide, T., Tsukahara, T., Ishikawa, M., & Maruyama, K. (2003). Measurement of depth dose distribution using plastic scintillator. Nippon Hoshasen Gijutsu Gakkai zasshi, 59(11), 1424-1431.

Measurement of depth dose distribution using plastic scintillator. / Hashimoto, Masatoshi; Kodama, Kiyoyuki; Hanada, Takashi; Ide, Tatsuya; Tsukahara, Tomoko; Ishikawa, Mitsuo; Maruyama, Koichi.

In: Nippon Hoshasen Gijutsu Gakkai zasshi, Vol. 59, No. 11, 2003, p. 1424-1431.

Research output: Contribution to journalArticle

Hashimoto, M, Kodama, K, Hanada, T, Ide, T, Tsukahara, T, Ishikawa, M & Maruyama, K 2003, 'Measurement of depth dose distribution using plastic scintillator', Nippon Hoshasen Gijutsu Gakkai zasshi, vol. 59, no. 11, pp. 1424-1431.
Hashimoto M, Kodama K, Hanada T, Ide T, Tsukahara T, Ishikawa M et al. Measurement of depth dose distribution using plastic scintillator. Nippon Hoshasen Gijutsu Gakkai zasshi. 2003;59(11):1424-1431.
Hashimoto, Masatoshi ; Kodama, Kiyoyuki ; Hanada, Takashi ; Ide, Tatsuya ; Tsukahara, Tomoko ; Ishikawa, Mitsuo ; Maruyama, Koichi. / Measurement of depth dose distribution using plastic scintillator. In: Nippon Hoshasen Gijutsu Gakkai zasshi. 2003 ; Vol. 59, No. 11. pp. 1424-1431.
@article{f18541cc1625427fbac7b6fc1bdae430,
title = "Measurement of depth dose distribution using plastic scintillator",
abstract = "We propose a new method to measure the depth dose distribution in matter. The method is to use organic plastic scintillator as a phantom, which has the effective atomic number close to that of human soft tissue, and to measure the position distribution of the scintillation light by a wavelength analyzer through a thread of plastic optical fiber. The purpose of this study is to examine possibility to measure the depth dose distribution with the proposed method. We carried out the following three measurements. First, we measured the effective field of view (FOV); the plastic optical fiber accepts the light from the FOV. Then, we measured the dose dependence of the amount of emission light. Finally, the light distribution in the depth direction by the present method was compared to the depth dose distribution by the ionization chamber. The results are as follows. The angular range of FOV is 7.03 degrees +/- 0.21 degrees The dose dependence shows a good linearity. In other words, the amount of emission light is proportional to the absorbed dose. Both of the light and the dose distributions are proportional each other, and they show the same attenuation in the region deeper than 15 mm. From these results, we conclude that the present method can be used for the measurement of the depth dose distribution in matter.",
author = "Masatoshi Hashimoto and Kiyoyuki Kodama and Takashi Hanada and Tatsuya Ide and Tomoko Tsukahara and Mitsuo Ishikawa and Koichi Maruyama",
year = "2003",
language = "English",
volume = "59",
pages = "1424--1431",
journal = "Nippon Hoshasen Gijutsu Gakkai zasshi",
issn = "0369-4305",
publisher = "Nihon Hoshasen Gijutsu Gakkai/Japanese Society of Radiological Technology",
number = "11",

}

TY - JOUR

T1 - Measurement of depth dose distribution using plastic scintillator

AU - Hashimoto, Masatoshi

AU - Kodama, Kiyoyuki

AU - Hanada, Takashi

AU - Ide, Tatsuya

AU - Tsukahara, Tomoko

AU - Ishikawa, Mitsuo

AU - Maruyama, Koichi

PY - 2003

Y1 - 2003

N2 - We propose a new method to measure the depth dose distribution in matter. The method is to use organic plastic scintillator as a phantom, which has the effective atomic number close to that of human soft tissue, and to measure the position distribution of the scintillation light by a wavelength analyzer through a thread of plastic optical fiber. The purpose of this study is to examine possibility to measure the depth dose distribution with the proposed method. We carried out the following three measurements. First, we measured the effective field of view (FOV); the plastic optical fiber accepts the light from the FOV. Then, we measured the dose dependence of the amount of emission light. Finally, the light distribution in the depth direction by the present method was compared to the depth dose distribution by the ionization chamber. The results are as follows. The angular range of FOV is 7.03 degrees +/- 0.21 degrees The dose dependence shows a good linearity. In other words, the amount of emission light is proportional to the absorbed dose. Both of the light and the dose distributions are proportional each other, and they show the same attenuation in the region deeper than 15 mm. From these results, we conclude that the present method can be used for the measurement of the depth dose distribution in matter.

AB - We propose a new method to measure the depth dose distribution in matter. The method is to use organic plastic scintillator as a phantom, which has the effective atomic number close to that of human soft tissue, and to measure the position distribution of the scintillation light by a wavelength analyzer through a thread of plastic optical fiber. The purpose of this study is to examine possibility to measure the depth dose distribution with the proposed method. We carried out the following three measurements. First, we measured the effective field of view (FOV); the plastic optical fiber accepts the light from the FOV. Then, we measured the dose dependence of the amount of emission light. Finally, the light distribution in the depth direction by the present method was compared to the depth dose distribution by the ionization chamber. The results are as follows. The angular range of FOV is 7.03 degrees +/- 0.21 degrees The dose dependence shows a good linearity. In other words, the amount of emission light is proportional to the absorbed dose. Both of the light and the dose distributions are proportional each other, and they show the same attenuation in the region deeper than 15 mm. From these results, we conclude that the present method can be used for the measurement of the depth dose distribution in matter.

UR - http://www.scopus.com/inward/record.url?scp=1642415852&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=1642415852&partnerID=8YFLogxK

M3 - Article

C2 - 14983130

AN - SCOPUS:1642415852

VL - 59

SP - 1424

EP - 1431

JO - Nippon Hoshasen Gijutsu Gakkai zasshi

JF - Nippon Hoshasen Gijutsu Gakkai zasshi

SN - 0369-4305

IS - 11

ER -