Near-wall motion of caged fluorescent dye in microchannel flows obtained from evanescent wave molecular tagging

Mitsuhisa Ichiyanagi, Yuriko Senga, Tsubasa Nakamura, Hiroki Fukumura, Yohei Sato

Research output: Contribution to journalArticle

Abstract

A molecular tagging technique utilizing evanescent wave illumination was developed to investigate the motion of a caged fluorescent dye in the vicinity of the microchannel wall surface in electroosmotic and pressure-driven flows. A line pattern in a buffer solution was written by a pulsed UV laser and the uncaged dye was excited by the evanescent wave with total internal reflection inside the glass wall using an objective lens. The velocities calculated by the measured displacement of the near-wall tagged region were compared with the results of molecular tagging using volume illumination, which represents the bulk flow information. Concerning electroosmotic flow, the micro-PIV technique using a confocal microscope system was applied to the microchannel rinsed by the caged fluorescein beforehand in comparison with a pure glass-PDMS microchannel to examine the effect of dye adsorption to the wall on the electroosmotic mobility. The electroosmotic mobility obtained by evanescent wave molecular tagging (EWMT) showed close to the micro-PIV measurement result near the glass wall for the rinsed case and the uncaged dye at the almost constant velocity remained in the depthwise illumination region. On the other hand, the dye velocity in pressure-driven flow by EWMT increased rapidly with respect to time. The uncaged dye convected to the streamwise direction dispersed toward the wall due to the concentration gradient of the dye, which was confirmed by the numerical simulations.

Original languageEnglish
Pages (from-to)2157-2169
Number of pages13
JournalNihon Kikai Gakkai Ronbunshu, B Hen/Transactions of the Japan Society of Mechanical Engineers, Part B
Volume77
Issue number783
Publication statusPublished - 2011
Externally publishedYes

Fingerprint

evanescent waves
microchannels
Microchannels
marking
Dyes
dyes
Lighting
illumination
particle image velocimetry
Glass
glass
information flow
ultraviolet lasers
Lenses
pulsed lasers
Microscopes
buffers
microscopes
lenses
Adsorption

Keywords

  • Adsorption
  • Caged fluorescent dye
  • Evanescent wave
  • Microchannel
  • Molecular tagging

ASJC Scopus subject areas

  • Mechanical Engineering
  • Condensed Matter Physics

Cite this

Near-wall motion of caged fluorescent dye in microchannel flows obtained from evanescent wave molecular tagging. / Ichiyanagi, Mitsuhisa; Senga, Yuriko; Nakamura, Tsubasa; Fukumura, Hiroki; Sato, Yohei.

In: Nihon Kikai Gakkai Ronbunshu, B Hen/Transactions of the Japan Society of Mechanical Engineers, Part B, Vol. 77, No. 783, 2011, p. 2157-2169.

Research output: Contribution to journalArticle

@article{4992c880ee094102a7782b2c28ba131b,
title = "Near-wall motion of caged fluorescent dye in microchannel flows obtained from evanescent wave molecular tagging",
abstract = "A molecular tagging technique utilizing evanescent wave illumination was developed to investigate the motion of a caged fluorescent dye in the vicinity of the microchannel wall surface in electroosmotic and pressure-driven flows. A line pattern in a buffer solution was written by a pulsed UV laser and the uncaged dye was excited by the evanescent wave with total internal reflection inside the glass wall using an objective lens. The velocities calculated by the measured displacement of the near-wall tagged region were compared with the results of molecular tagging using volume illumination, which represents the bulk flow information. Concerning electroosmotic flow, the micro-PIV technique using a confocal microscope system was applied to the microchannel rinsed by the caged fluorescein beforehand in comparison with a pure glass-PDMS microchannel to examine the effect of dye adsorption to the wall on the electroosmotic mobility. The electroosmotic mobility obtained by evanescent wave molecular tagging (EWMT) showed close to the micro-PIV measurement result near the glass wall for the rinsed case and the uncaged dye at the almost constant velocity remained in the depthwise illumination region. On the other hand, the dye velocity in pressure-driven flow by EWMT increased rapidly with respect to time. The uncaged dye convected to the streamwise direction dispersed toward the wall due to the concentration gradient of the dye, which was confirmed by the numerical simulations.",
keywords = "Adsorption, Caged fluorescent dye, Evanescent wave, Microchannel, Molecular tagging",
author = "Mitsuhisa Ichiyanagi and Yuriko Senga and Tsubasa Nakamura and Hiroki Fukumura and Yohei Sato",
year = "2011",
language = "English",
volume = "77",
pages = "2157--2169",
journal = "Nihon Kikai Gakkai Ronbunshu, B Hen/Transactions of the Japan Society of Mechanical Engineers, Part B",
issn = "0387-5016",
publisher = "Japan Society of Mechanical Engineers",
number = "783",

}

TY - JOUR

T1 - Near-wall motion of caged fluorescent dye in microchannel flows obtained from evanescent wave molecular tagging

AU - Ichiyanagi, Mitsuhisa

AU - Senga, Yuriko

AU - Nakamura, Tsubasa

AU - Fukumura, Hiroki

AU - Sato, Yohei

PY - 2011

Y1 - 2011

N2 - A molecular tagging technique utilizing evanescent wave illumination was developed to investigate the motion of a caged fluorescent dye in the vicinity of the microchannel wall surface in electroosmotic and pressure-driven flows. A line pattern in a buffer solution was written by a pulsed UV laser and the uncaged dye was excited by the evanescent wave with total internal reflection inside the glass wall using an objective lens. The velocities calculated by the measured displacement of the near-wall tagged region were compared with the results of molecular tagging using volume illumination, which represents the bulk flow information. Concerning electroosmotic flow, the micro-PIV technique using a confocal microscope system was applied to the microchannel rinsed by the caged fluorescein beforehand in comparison with a pure glass-PDMS microchannel to examine the effect of dye adsorption to the wall on the electroosmotic mobility. The electroosmotic mobility obtained by evanescent wave molecular tagging (EWMT) showed close to the micro-PIV measurement result near the glass wall for the rinsed case and the uncaged dye at the almost constant velocity remained in the depthwise illumination region. On the other hand, the dye velocity in pressure-driven flow by EWMT increased rapidly with respect to time. The uncaged dye convected to the streamwise direction dispersed toward the wall due to the concentration gradient of the dye, which was confirmed by the numerical simulations.

AB - A molecular tagging technique utilizing evanescent wave illumination was developed to investigate the motion of a caged fluorescent dye in the vicinity of the microchannel wall surface in electroosmotic and pressure-driven flows. A line pattern in a buffer solution was written by a pulsed UV laser and the uncaged dye was excited by the evanescent wave with total internal reflection inside the glass wall using an objective lens. The velocities calculated by the measured displacement of the near-wall tagged region were compared with the results of molecular tagging using volume illumination, which represents the bulk flow information. Concerning electroosmotic flow, the micro-PIV technique using a confocal microscope system was applied to the microchannel rinsed by the caged fluorescein beforehand in comparison with a pure glass-PDMS microchannel to examine the effect of dye adsorption to the wall on the electroosmotic mobility. The electroosmotic mobility obtained by evanescent wave molecular tagging (EWMT) showed close to the micro-PIV measurement result near the glass wall for the rinsed case and the uncaged dye at the almost constant velocity remained in the depthwise illumination region. On the other hand, the dye velocity in pressure-driven flow by EWMT increased rapidly with respect to time. The uncaged dye convected to the streamwise direction dispersed toward the wall due to the concentration gradient of the dye, which was confirmed by the numerical simulations.

KW - Adsorption

KW - Caged fluorescent dye

KW - Evanescent wave

KW - Microchannel

KW - Molecular tagging

UR - http://www.scopus.com/inward/record.url?scp=84859618238&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84859618238&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:84859618238

VL - 77

SP - 2157

EP - 2169

JO - Nihon Kikai Gakkai Ronbunshu, B Hen/Transactions of the Japan Society of Mechanical Engineers, Part B

JF - Nihon Kikai Gakkai Ronbunshu, B Hen/Transactions of the Japan Society of Mechanical Engineers, Part B

SN - 0387-5016

IS - 783

ER -