Nondestructive evaluation of blood flow in a dialyzer using X-ray computed tomography

Yoshitada Sakai, Shigehisa Wada, Hiroshi Matsumoto, Tomoko Suyama, Osamu Ohno, Izumi Anno

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

It is very important to observe the concentrations and flow patterns of blood through a dialyzer to evaluate its function and to obtain the most appropriate design. We established a visualization method for the blood flow pattern in a dialyzer using X-ray computed tomography, and investigated the so-called internal filtration phenomenon. The results obtained were as follows: (1) The influence of 5% BaSO4, which was added to the blood as a contrast medium, on the filtration rate of the dialyzer was minimal. (2) The relationship between the concentration of BaSO4 and the Hounsfield unit value was expressed by linear regression. (3) Hounsfield unit values increased massively just after blood entered the dialyzer and peak values increased with dialysate perfusion under the following conditions: the dialyzer (BS-1.6UL, polysulfone hollow fibers) was used, and bovine blood with 5% BaSO4 added was used at a blood flow rate of 200 ml/min. The dialysate flow rate was 500 ml/ min and the slice thickness of X-ray computed tomography was 1-10 mm. (4) It was observed that blood flowed slightly faster in the center than the peripheral portion of the dialyzer, when the flow pattern was followed after pulse injection of blood containing 20% BaSO 4 into the dialyzer. It was concluded that this method could possibly be utilized not only qualitatively but also quantitatively for observation of the real state of blood flow and in designing dialyzers.

Original languageEnglish
Pages (from-to)197-204
Number of pages8
JournalJournal of Artificial Organs
Volume6
Issue number3
DOIs
Publication statusPublished - 2003

Fingerprint

X Ray Computed Tomography
Tomography
Blood
X rays
Flow patterns
Dialysis Solutions
Artificial Kidneys
Flow rate
Contrast media
Polysulfones
Contrast Media
Linear regression
Linear Models
Perfusion
Observation
Visualization
Injections
Fibers

Keywords

  • BaSO
  • Dialyzer
  • Flow pattern
  • Internal filtration
  • X-ray CT

ASJC Scopus subject areas

  • Biophysics

Cite this

Nondestructive evaluation of blood flow in a dialyzer using X-ray computed tomography. / Sakai, Yoshitada; Wada, Shigehisa; Matsumoto, Hiroshi; Suyama, Tomoko; Ohno, Osamu; Anno, Izumi.

In: Journal of Artificial Organs, Vol. 6, No. 3, 2003, p. 197-204.

Research output: Contribution to journalArticle

Sakai, Yoshitada ; Wada, Shigehisa ; Matsumoto, Hiroshi ; Suyama, Tomoko ; Ohno, Osamu ; Anno, Izumi. / Nondestructive evaluation of blood flow in a dialyzer using X-ray computed tomography. In: Journal of Artificial Organs. 2003 ; Vol. 6, No. 3. pp. 197-204.
@article{8f33a84923d24ceba44e8425b4513f64,
title = "Nondestructive evaluation of blood flow in a dialyzer using X-ray computed tomography",
abstract = "It is very important to observe the concentrations and flow patterns of blood through a dialyzer to evaluate its function and to obtain the most appropriate design. We established a visualization method for the blood flow pattern in a dialyzer using X-ray computed tomography, and investigated the so-called internal filtration phenomenon. The results obtained were as follows: (1) The influence of 5{\%} BaSO4, which was added to the blood as a contrast medium, on the filtration rate of the dialyzer was minimal. (2) The relationship between the concentration of BaSO4 and the Hounsfield unit value was expressed by linear regression. (3) Hounsfield unit values increased massively just after blood entered the dialyzer and peak values increased with dialysate perfusion under the following conditions: the dialyzer (BS-1.6UL, polysulfone hollow fibers) was used, and bovine blood with 5{\%} BaSO4 added was used at a blood flow rate of 200 ml/min. The dialysate flow rate was 500 ml/ min and the slice thickness of X-ray computed tomography was 1-10 mm. (4) It was observed that blood flowed slightly faster in the center than the peripheral portion of the dialyzer, when the flow pattern was followed after pulse injection of blood containing 20{\%} BaSO 4 into the dialyzer. It was concluded that this method could possibly be utilized not only qualitatively but also quantitatively for observation of the real state of blood flow and in designing dialyzers.",
keywords = "BaSO, Dialyzer, Flow pattern, Internal filtration, X-ray CT",
author = "Yoshitada Sakai and Shigehisa Wada and Hiroshi Matsumoto and Tomoko Suyama and Osamu Ohno and Izumi Anno",
year = "2003",
doi = "10.1007/s10047-003-0228-6",
language = "English",
volume = "6",
pages = "197--204",
journal = "Journal of Artificial Organs",
issn = "1434-7229",
publisher = "Springer Japan",
number = "3",

}

TY - JOUR

T1 - Nondestructive evaluation of blood flow in a dialyzer using X-ray computed tomography

AU - Sakai, Yoshitada

AU - Wada, Shigehisa

AU - Matsumoto, Hiroshi

AU - Suyama, Tomoko

AU - Ohno, Osamu

AU - Anno, Izumi

PY - 2003

Y1 - 2003

N2 - It is very important to observe the concentrations and flow patterns of blood through a dialyzer to evaluate its function and to obtain the most appropriate design. We established a visualization method for the blood flow pattern in a dialyzer using X-ray computed tomography, and investigated the so-called internal filtration phenomenon. The results obtained were as follows: (1) The influence of 5% BaSO4, which was added to the blood as a contrast medium, on the filtration rate of the dialyzer was minimal. (2) The relationship between the concentration of BaSO4 and the Hounsfield unit value was expressed by linear regression. (3) Hounsfield unit values increased massively just after blood entered the dialyzer and peak values increased with dialysate perfusion under the following conditions: the dialyzer (BS-1.6UL, polysulfone hollow fibers) was used, and bovine blood with 5% BaSO4 added was used at a blood flow rate of 200 ml/min. The dialysate flow rate was 500 ml/ min and the slice thickness of X-ray computed tomography was 1-10 mm. (4) It was observed that blood flowed slightly faster in the center than the peripheral portion of the dialyzer, when the flow pattern was followed after pulse injection of blood containing 20% BaSO 4 into the dialyzer. It was concluded that this method could possibly be utilized not only qualitatively but also quantitatively for observation of the real state of blood flow and in designing dialyzers.

AB - It is very important to observe the concentrations and flow patterns of blood through a dialyzer to evaluate its function and to obtain the most appropriate design. We established a visualization method for the blood flow pattern in a dialyzer using X-ray computed tomography, and investigated the so-called internal filtration phenomenon. The results obtained were as follows: (1) The influence of 5% BaSO4, which was added to the blood as a contrast medium, on the filtration rate of the dialyzer was minimal. (2) The relationship between the concentration of BaSO4 and the Hounsfield unit value was expressed by linear regression. (3) Hounsfield unit values increased massively just after blood entered the dialyzer and peak values increased with dialysate perfusion under the following conditions: the dialyzer (BS-1.6UL, polysulfone hollow fibers) was used, and bovine blood with 5% BaSO4 added was used at a blood flow rate of 200 ml/min. The dialysate flow rate was 500 ml/ min and the slice thickness of X-ray computed tomography was 1-10 mm. (4) It was observed that blood flowed slightly faster in the center than the peripheral portion of the dialyzer, when the flow pattern was followed after pulse injection of blood containing 20% BaSO 4 into the dialyzer. It was concluded that this method could possibly be utilized not only qualitatively but also quantitatively for observation of the real state of blood flow and in designing dialyzers.

KW - BaSO

KW - Dialyzer

KW - Flow pattern

KW - Internal filtration

KW - X-ray CT

UR - http://www.scopus.com/inward/record.url?scp=0141891979&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0141891979&partnerID=8YFLogxK

U2 - 10.1007/s10047-003-0228-6

DO - 10.1007/s10047-003-0228-6

M3 - Article

VL - 6

SP - 197

EP - 204

JO - Journal of Artificial Organs

JF - Journal of Artificial Organs

SN - 1434-7229

IS - 3

ER -