Abstract

Field inhomogeneity related phase errors in multi-shot echo planar imaging (EPI) are directly visualized and analyzed in the spatial frequency domain data or 'k-space'. The echo time shift (ETS) technique incrementally moves the position of the echo train and improves the phase error function by redistributing phase discontinuities away from the center of k-space.

Original languageEnglish
Pages (from-to)535-539
Number of pages5
JournalMagnetic Resonance in Medicine
Volume32
Issue number4
DOIs
Publication statusPublished - 1994
Externally publishedYes

Fingerprint

Echo-Planar Imaging

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Phase errors in multi-shot echo planar imaging. / Feinberg, D. A.; Oshio, Koichi.

In: Magnetic Resonance in Medicine, Vol. 32, No. 4, 1994, p. 535-539.

Research output: Contribution to journalArticle

Feinberg, D. A. ; Oshio, Koichi. / Phase errors in multi-shot echo planar imaging. In: Magnetic Resonance in Medicine. 1994 ; Vol. 32, No. 4. pp. 535-539.
@article{81639c9059c7468c8d89604aa340e632,
title = "Phase errors in multi-shot echo planar imaging",
abstract = "Field inhomogeneity related phase errors in multi-shot echo planar imaging (EPI) are directly visualized and analyzed in the spatial frequency domain data or 'k-space'. The echo time shift (ETS) technique incrementally moves the position of the echo train and improves the phase error function by redistributing phase discontinuities away from the center of k-space.",
author = "Feinberg, {D. A.} and Koichi Oshio",
year = "1994",
doi = "10.1002/mrm.1910320418",
language = "English",
volume = "32",
pages = "535--539",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "4",

}

TY - JOUR

T1 - Phase errors in multi-shot echo planar imaging

AU - Feinberg, D. A.

AU - Oshio, Koichi

PY - 1994

Y1 - 1994

N2 - Field inhomogeneity related phase errors in multi-shot echo planar imaging (EPI) are directly visualized and analyzed in the spatial frequency domain data or 'k-space'. The echo time shift (ETS) technique incrementally moves the position of the echo train and improves the phase error function by redistributing phase discontinuities away from the center of k-space.

AB - Field inhomogeneity related phase errors in multi-shot echo planar imaging (EPI) are directly visualized and analyzed in the spatial frequency domain data or 'k-space'. The echo time shift (ETS) technique incrementally moves the position of the echo train and improves the phase error function by redistributing phase discontinuities away from the center of k-space.

UR - http://www.scopus.com/inward/record.url?scp=0028145655&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0028145655&partnerID=8YFLogxK

U2 - 10.1002/mrm.1910320418

DO - 10.1002/mrm.1910320418

M3 - Article

VL - 32

SP - 535

EP - 539

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 4

ER -