Silent navigator-triggered silent MRI of the abdomen

Yuji Iwadate, Atsushi Nozaki, Yoshinobu Nunokawa, Shigeo Okuda, Masahiro Jinzaki, Hiroyuki Kabasawa

Research output: Contribution to journalArticle

Abstract

Purpose: To develop and demonstrate the feasibility of a silent respiratory navigator technique for prospective triggering, which was incorporated into a three-dimensional radial zero-echo-time sequence for respiratory navigated silent abdominal imaging. Methods: A nonselective hard excitation radiofrequency pulse was used for the navigator sequence with a derated readout gradient, to avoid generation of high levels of acoustic noise. The acquired navigator signals were processed in real time and used for prospective triggering of the zero-echo-time sequence. Ten healthy volunteers were scanned using the proposed and conventional techniques at 1.5T. An acoustic noise measurement with A-weighted continuous equivalent sound pressure level was also performed. Results: The sound pressure-level values of the background noise, zero-echo-time imaging, conventional, and silent navigators were 68.3, 68.4, 102.5, and 69.4 dBA, respectively. Excellent correlation with correlation coefficients greater than 0.9 was observed between the bellows signals and displacement values calculated from the navigators. Sharpness of the portal vein of both conventional and silent navigator-triggered images was significantly higher than those of nontriggered images. Conclusions: The silent navigator-triggered zero-echo-time technique is feasible and might improve image quality and workflow of abdominal MRI of patients who are prone to acoustic noise.

Original languageEnglish
JournalMagnetic Resonance in Medicine
DOIs
Publication statusAccepted/In press - 2017

Fingerprint

Abdomen
Noise
Acoustics
Pressure
Workflow
Portal Vein
Healthy Volunteers

Keywords

  • Abdominal imaging
  • Navigator
  • Respiratory motion
  • Silent MRI

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Silent navigator-triggered silent MRI of the abdomen. / Iwadate, Yuji; Nozaki, Atsushi; Nunokawa, Yoshinobu; Okuda, Shigeo; Jinzaki, Masahiro; Kabasawa, Hiroyuki.

In: Magnetic Resonance in Medicine, 2017.

Research output: Contribution to journalArticle

Iwadate, Yuji ; Nozaki, Atsushi ; Nunokawa, Yoshinobu ; Okuda, Shigeo ; Jinzaki, Masahiro ; Kabasawa, Hiroyuki. / Silent navigator-triggered silent MRI of the abdomen. In: Magnetic Resonance in Medicine. 2017.
@article{1afb067f6e544f6bbced9e835a0bdcd5,
title = "Silent navigator-triggered silent MRI of the abdomen",
abstract = "Purpose: To develop and demonstrate the feasibility of a silent respiratory navigator technique for prospective triggering, which was incorporated into a three-dimensional radial zero-echo-time sequence for respiratory navigated silent abdominal imaging. Methods: A nonselective hard excitation radiofrequency pulse was used for the navigator sequence with a derated readout gradient, to avoid generation of high levels of acoustic noise. The acquired navigator signals were processed in real time and used for prospective triggering of the zero-echo-time sequence. Ten healthy volunteers were scanned using the proposed and conventional techniques at 1.5T. An acoustic noise measurement with A-weighted continuous equivalent sound pressure level was also performed. Results: The sound pressure-level values of the background noise, zero-echo-time imaging, conventional, and silent navigators were 68.3, 68.4, 102.5, and 69.4 dBA, respectively. Excellent correlation with correlation coefficients greater than 0.9 was observed between the bellows signals and displacement values calculated from the navigators. Sharpness of the portal vein of both conventional and silent navigator-triggered images was significantly higher than those of nontriggered images. Conclusions: The silent navigator-triggered zero-echo-time technique is feasible and might improve image quality and workflow of abdominal MRI of patients who are prone to acoustic noise.",
keywords = "Abdominal imaging, Navigator, Respiratory motion, Silent MRI",
author = "Yuji Iwadate and Atsushi Nozaki and Yoshinobu Nunokawa and Shigeo Okuda and Masahiro Jinzaki and Hiroyuki Kabasawa",
year = "2017",
doi = "10.1002/mrm.26869",
language = "English",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",

}

TY - JOUR

T1 - Silent navigator-triggered silent MRI of the abdomen

AU - Iwadate, Yuji

AU - Nozaki, Atsushi

AU - Nunokawa, Yoshinobu

AU - Okuda, Shigeo

AU - Jinzaki, Masahiro

AU - Kabasawa, Hiroyuki

PY - 2017

Y1 - 2017

N2 - Purpose: To develop and demonstrate the feasibility of a silent respiratory navigator technique for prospective triggering, which was incorporated into a three-dimensional radial zero-echo-time sequence for respiratory navigated silent abdominal imaging. Methods: A nonselective hard excitation radiofrequency pulse was used for the navigator sequence with a derated readout gradient, to avoid generation of high levels of acoustic noise. The acquired navigator signals were processed in real time and used for prospective triggering of the zero-echo-time sequence. Ten healthy volunteers were scanned using the proposed and conventional techniques at 1.5T. An acoustic noise measurement with A-weighted continuous equivalent sound pressure level was also performed. Results: The sound pressure-level values of the background noise, zero-echo-time imaging, conventional, and silent navigators were 68.3, 68.4, 102.5, and 69.4 dBA, respectively. Excellent correlation with correlation coefficients greater than 0.9 was observed between the bellows signals and displacement values calculated from the navigators. Sharpness of the portal vein of both conventional and silent navigator-triggered images was significantly higher than those of nontriggered images. Conclusions: The silent navigator-triggered zero-echo-time technique is feasible and might improve image quality and workflow of abdominal MRI of patients who are prone to acoustic noise.

AB - Purpose: To develop and demonstrate the feasibility of a silent respiratory navigator technique for prospective triggering, which was incorporated into a three-dimensional radial zero-echo-time sequence for respiratory navigated silent abdominal imaging. Methods: A nonselective hard excitation radiofrequency pulse was used for the navigator sequence with a derated readout gradient, to avoid generation of high levels of acoustic noise. The acquired navigator signals were processed in real time and used for prospective triggering of the zero-echo-time sequence. Ten healthy volunteers were scanned using the proposed and conventional techniques at 1.5T. An acoustic noise measurement with A-weighted continuous equivalent sound pressure level was also performed. Results: The sound pressure-level values of the background noise, zero-echo-time imaging, conventional, and silent navigators were 68.3, 68.4, 102.5, and 69.4 dBA, respectively. Excellent correlation with correlation coefficients greater than 0.9 was observed between the bellows signals and displacement values calculated from the navigators. Sharpness of the portal vein of both conventional and silent navigator-triggered images was significantly higher than those of nontriggered images. Conclusions: The silent navigator-triggered zero-echo-time technique is feasible and might improve image quality and workflow of abdominal MRI of patients who are prone to acoustic noise.

KW - Abdominal imaging

KW - Navigator

KW - Respiratory motion

KW - Silent MRI

UR - http://www.scopus.com/inward/record.url?scp=85027194738&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85027194738&partnerID=8YFLogxK

U2 - 10.1002/mrm.26869

DO - 10.1002/mrm.26869

M3 - Article

C2 - 28782173

AN - SCOPUS:85027194738

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

ER -