T1 - The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm
AU - Ishihara, Hironori
AU - Sugo, Yoshihiro
AU - Tsutsui, Masato
AU - Yamada, Takashige
AU - Sato, Tetsufumi
AU - Akazawa, Toshimasa
AU - Sato, Nobukazu
AU - Yamashita, Koichi
AU - Takeda, Junzo
N1 - Funding Information:
Acknowledgments Supported by equipment provided by Nihon Kohden Corporation, Japan.
PY - 2012/12
Y1 - 2012/12
N2 - A new non-invasive continuous cardiac output (esCCO) monitoring system solely utilizing a routine cardiovascular monitor was developed, even though a reference cardiac output (CO) is consistently required. Subsequently, a non-invasive patient information CO calibration together with a new automated exclusion algorithm was implemented in the esCCO system. We evaluated the accuracy and trending ability of the new esCCO system. Either operative or postoperative data of a multicenter study in Japan for evaluation of the accuracy of the original version of esCCO system were used to develop the new esCCO system. A total of 207 patients, mostly cardiac surgical patients, were enrolled in the study. Data were manually reviewed to formulate a new automated exclusion algorithm with enhanced accuracy. Then, a new esCCO system based on a patient information calibration together with the automated exclusion algorithm was developed. CO measured with a new esCCO system was compared with the corresponding intermittent bolus thermodilution CO (ICO) utilizing statistical methods including polar plots analysis. A total of 465 sets of CO data obtained using the new esCCO system were evaluated. The difference in the CO value between the new esCCO and ICO was 0.34 ± 1.50 (SD) L/min (95 % confidence limits of -2.60 to 3.28 L/min). The percentage error was 69.6 %. Polar plots analysis showed that the mean polar angle was -1.6 and radial limits of agreement were ±53.3. This study demonstrates that the patient information calibration is clinically useful as ICO, but trending ability of the new esCCO system is not clinically acceptable as judged by percentage error and polar plots analysis, even though it's trending ability is comparable with currently available arterial waveform analysis methods.
AB - A new non-invasive continuous cardiac output (esCCO) monitoring system solely utilizing a routine cardiovascular monitor was developed, even though a reference cardiac output (CO) is consistently required. Subsequently, a non-invasive patient information CO calibration together with a new automated exclusion algorithm was implemented in the esCCO system. We evaluated the accuracy and trending ability of the new esCCO system. Either operative or postoperative data of a multicenter study in Japan for evaluation of the accuracy of the original version of esCCO system were used to develop the new esCCO system. A total of 207 patients, mostly cardiac surgical patients, were enrolled in the study. Data were manually reviewed to formulate a new automated exclusion algorithm with enhanced accuracy. Then, a new esCCO system based on a patient information calibration together with the automated exclusion algorithm was developed. CO measured with a new esCCO system was compared with the corresponding intermittent bolus thermodilution CO (ICO) utilizing statistical methods including polar plots analysis. A total of 465 sets of CO data obtained using the new esCCO system were evaluated. The difference in the CO value between the new esCCO and ICO was 0.34 ± 1.50 (SD) L/min (95 % confidence limits of -2.60 to 3.28 L/min). The percentage error was 69.6 %. Polar plots analysis showed that the mean polar angle was -1.6 and radial limits of agreement were ±53.3. This study demonstrates that the patient information calibration is clinically useful as ICO, but trending ability of the new esCCO system is not clinically acceptable as judged by percentage error and polar plots analysis, even though it's trending ability is comparable with currently available arterial waveform analysis methods.
KW - Cardiac output
KW - Measurement technique
KW - Non-invasive calibration
KW - Pulse contour analysis
KW - Pulse wave transit time
UR - http://www.scopus.com/inward/record.url?scp=84877106251&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84877106251&partnerID=8YFLogxK
U2 - 10.1007/s10877-012-9384-7
DO - 10.1007/s10877-012-9384-7
M3 - Article
C2 - 22854918
AN - SCOPUS:84877106251
SN - 1387-1307
VL - 26
SP - 465
EP - 471
JO - Journal of Clinical Monitoring and Computing
JF - Journal of Clinical Monitoring and Computing
IS - 6
ER -