The reproducibility of a HeadFix relocatable fixation system

Analysis using the stereotactic coordinates of bilateral incus and the top of the crista galli obtained from a serial CT scan

Etsuo Kunieda, Yohei Oku, Junichi Fukada, Osamu Kawaguchi, Hideyuki Shiba, Atsuya Takeda, Atsushi Kubo

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

We analysed the repositioning accuracy of bite-plate fixations from serial QA-CT (quality-assurance CT) taken during the course of stereotactic radiotherapy. A total of 72 series of CT examinations from 15 consecutive patients, who underwent stereotactic radiotherapy for various intracranial tumours, were examined. Three or four CT scans were obtained for the purpose of QA for the right and left incus, as well as the crista galli. The stereotactic coordinates of the centres of the incus and the top of the crista galli were semi-automatically obtained for each QA-CT scan. Positional displacements for these anatomical reference points and the centre of the points were obtained. The mean displacements for these points in the 3D directions ranged from -0.10 to 0.08 mm (standard deviations: 0.44-0.94). The absolute positional displacement ranged from 0.93 to 1.09 mm (standard deviations: 0.52-0.88 mm). The rotations of the head were 0.49 0.36°. Our 3D measurement technique using anatomical landmarks revealed excellent stability of the mouthpiece fixation system in terms of translational and rotational displacements. This technique can also be used as a QA method for other fixation devices.

Original languageEnglish
JournalPhysics in Medicine and Biology
Volume54
Issue number10
DOIs
Publication statusPublished - 2009

Fingerprint

Incus
Systems Analysis
Radiotherapy
Bites and Stings
Head
Equipment and Supplies
Neoplasms
Direction compound

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

@article{cbf33185b13f4f42ae4af47b73dd7888,
title = "The reproducibility of a HeadFix relocatable fixation system: Analysis using the stereotactic coordinates of bilateral incus and the top of the crista galli obtained from a serial CT scan",
abstract = "We analysed the repositioning accuracy of bite-plate fixations from serial QA-CT (quality-assurance CT) taken during the course of stereotactic radiotherapy. A total of 72 series of CT examinations from 15 consecutive patients, who underwent stereotactic radiotherapy for various intracranial tumours, were examined. Three or four CT scans were obtained for the purpose of QA for the right and left incus, as well as the crista galli. The stereotactic coordinates of the centres of the incus and the top of the crista galli were semi-automatically obtained for each QA-CT scan. Positional displacements for these anatomical reference points and the centre of the points were obtained. The mean displacements for these points in the 3D directions ranged from -0.10 to 0.08 mm (standard deviations: 0.44-0.94). The absolute positional displacement ranged from 0.93 to 1.09 mm (standard deviations: 0.52-0.88 mm). The rotations of the head were 0.49 0.36°. Our 3D measurement technique using anatomical landmarks revealed excellent stability of the mouthpiece fixation system in terms of translational and rotational displacements. This technique can also be used as a QA method for other fixation devices.",
author = "Etsuo Kunieda and Yohei Oku and Junichi Fukada and Osamu Kawaguchi and Hideyuki Shiba and Atsuya Takeda and Atsushi Kubo",
year = "2009",
doi = "10.1088/0031-9155/54/10/N03",
language = "English",
volume = "54",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "10",

}

TY - JOUR

T1 - The reproducibility of a HeadFix relocatable fixation system

T2 - Analysis using the stereotactic coordinates of bilateral incus and the top of the crista galli obtained from a serial CT scan

AU - Kunieda, Etsuo

AU - Oku, Yohei

AU - Fukada, Junichi

AU - Kawaguchi, Osamu

AU - Shiba, Hideyuki

AU - Takeda, Atsuya

AU - Kubo, Atsushi

PY - 2009

Y1 - 2009

N2 - We analysed the repositioning accuracy of bite-plate fixations from serial QA-CT (quality-assurance CT) taken during the course of stereotactic radiotherapy. A total of 72 series of CT examinations from 15 consecutive patients, who underwent stereotactic radiotherapy for various intracranial tumours, were examined. Three or four CT scans were obtained for the purpose of QA for the right and left incus, as well as the crista galli. The stereotactic coordinates of the centres of the incus and the top of the crista galli were semi-automatically obtained for each QA-CT scan. Positional displacements for these anatomical reference points and the centre of the points were obtained. The mean displacements for these points in the 3D directions ranged from -0.10 to 0.08 mm (standard deviations: 0.44-0.94). The absolute positional displacement ranged from 0.93 to 1.09 mm (standard deviations: 0.52-0.88 mm). The rotations of the head were 0.49 0.36°. Our 3D measurement technique using anatomical landmarks revealed excellent stability of the mouthpiece fixation system in terms of translational and rotational displacements. This technique can also be used as a QA method for other fixation devices.

AB - We analysed the repositioning accuracy of bite-plate fixations from serial QA-CT (quality-assurance CT) taken during the course of stereotactic radiotherapy. A total of 72 series of CT examinations from 15 consecutive patients, who underwent stereotactic radiotherapy for various intracranial tumours, were examined. Three or four CT scans were obtained for the purpose of QA for the right and left incus, as well as the crista galli. The stereotactic coordinates of the centres of the incus and the top of the crista galli were semi-automatically obtained for each QA-CT scan. Positional displacements for these anatomical reference points and the centre of the points were obtained. The mean displacements for these points in the 3D directions ranged from -0.10 to 0.08 mm (standard deviations: 0.44-0.94). The absolute positional displacement ranged from 0.93 to 1.09 mm (standard deviations: 0.52-0.88 mm). The rotations of the head were 0.49 0.36°. Our 3D measurement technique using anatomical landmarks revealed excellent stability of the mouthpiece fixation system in terms of translational and rotational displacements. This technique can also be used as a QA method for other fixation devices.

UR - http://www.scopus.com/inward/record.url?scp=67650720997&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=67650720997&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/54/10/N03

DO - 10.1088/0031-9155/54/10/N03

M3 - Article

VL - 54

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 10

ER -