Application of real-time radiation dosimetry using a new silicon LET sensor

Tadayoshi Doke, Takayoshi Hayashi, Jun Kikuchi, Shunji Nagaoka, T. Nakano, T. Sakaguchi, K. Terasawa, G. D. Badhwar

研究成果: Article査読

3 被引用数 (Scopus)

抄録

A new type of real-time radiation monitoring device, RRMD-III, consisting of three double-sided silicon strip detectors (DSSDs), has been developed and tested on-board the Space Shuttle mission STS-84. The test succeeded in measuring the linear energy transfer (LET) distribution over the range of 0.2 keV/μm to 600 keV/μm for 178 h. The Shuttle cruised at an altitude of 300 to 400 km and an inclination angle of 51.6°for 221.3 h, which is equivalent to the International Space Station orbit. The LET distribution obtained for particles was investigated by separating it into galactic cosmic ray (GCR) particles and trapped particles in the South Atlantic Anomaly (SAA) region. The result shows that the contribution in dose-equivalent due to GCR particles is almost equal to that from trapped particles. The total absorbed dose rate during the mission was 0.611 mGy/day; the effective quality factor, 1.64; and the dose equivalent rate, 0.998 mSv/day. The average absorbed dose rates are 0.158 mGy/min for GCR particles and 3.67 mGy/min for trapped particles. The effective quality factors are 2.48 for GCR particles and 1.19 for trapped particles. The absorbed doses obtained by the RRMD-III and a conventional method using TLD (Mg2SiO4), which was placed around the RRMD-III were compared. It was found that the TLDs showed a lower efficiency, just 58% of absorbed dose registered by the RRMD-III. Copyright (C) 1999 Elsevier Science B.V.

本文言語English
ページ(範囲)191-202
ページ数12
ジャーナルMutation Research - Fundamental and Molecular Mechanisms of Mutagenesis
430
2
DOI
出版ステータスPublished - 1999 12月 6
外部発表はい

ASJC Scopus subject areas

  • 分子生物学
  • 遺伝学
  • 健康、毒物学および変異誘発

フィンガープリント

「Application of real-time radiation dosimetry using a new silicon LET sensor」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル