Cds nanoparticle-based biosensor development for aflatoxin determination

Moh Hayat, Endang Saepudin, Yasuaki Einaga, Tribidasari A. Ivandini

研究成果: Article査読

1 被引用数 (Scopus)

抄録

An anodic stripping voltammetry detection method with lateral flow immunochromatographic assay for determining aflatoxin M1 was developed. Cadmium sulfide nanoparticles (CdSNPs) were employed as a label for aflatoxin antibody. Conjugation between the CdSNPs and aflatoxin antibody was performed as the first step. The CdSNP-antibody conjugate was then immobilized on a test strip on a conjugate pad. When the conjugate interacted with the sample containing aflatoxin, it was captured by the aflatoxin in the test zone. The CdSNPs retained in the test zone were then measured by anodic stripping voltammetry using a boron-doped diamond electrode. A linear concentration range of 0-70 ppb aflatoxin, with a sensitivity of 20 μA/mM and a detection limit of 30 ppb, was achieved by this method.

本文言語English
ページ(範囲)787-797
ページ数11
ジャーナルInternational Journal of Technology
10
4
DOI
出版ステータスPublished - 2019 7

ASJC Scopus subject areas

  • Engineering(all)
  • Strategy and Management
  • Management of Technology and Innovation

フィンガープリント 「Cds nanoparticle-based biosensor development for aflatoxin determination」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。