Clinical usefulness of digital radiography in the gastrointestinal tract: Efficacy of magnification method

Yoshinori Sugino, Yutaka Imai, Hirohisa Fujisawa, Kyouichi Hiramatsu, Hiroshi Amoh, Kenji Kumakura

研究成果: Article

4 引用 (Scopus)

抜粋

We assessed the performance capabilities of image intensifier digital radiography (II DR) in the detection of minute lesions in patients with early stomach cancer. The DR system was a prototype II DR system developed by Toshiba Corp (Tokyo, Japan). This system was able to acquire images with a 1,024-× 1,024-pixel matrix and 13 bits. Radiography was performed using a 0.3-mm tube focus. For the detectability of early stomach cancer, DR was judged to be superior to conventional screen-film system (CFSS) (DR superior, 55.7%; CFSS superior, 22.6%). In depicting the characteristics of the surface of the lesion, DR was also judged to be superior to CFSS (DR superior, 56.6%; CFSS superior, 17.0%). The II DR system used in this study was able to achieve almost the same spatial resolution as conventional radiography using the magnification method. It was also able to visualize subtle findings of early gastric cancer more clearly by the use of postprocessing. In addition, II DR has the advantages of reducing the patient exposure dose and permitting the acquisition of real-time images.

元の言語English
ページ(範囲)84-88
ページ数5
ジャーナルJournal of Digital Imaging
8
発行部数1 Supplement
DOI
出版物ステータスPublished - 1995 2 1

    フィンガープリント

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging
  • Computer Science Applications

これを引用

Sugino, Y., Imai, Y., Fujisawa, H., Hiramatsu, K., Amoh, H., & Kumakura, K. (1995). Clinical usefulness of digital radiography in the gastrointestinal tract: Efficacy of magnification method. Journal of Digital Imaging, 8(1 Supplement), 84-88. https://doi.org/10.1007/BF03168074