TY - JOUR
T1 - Contact-free unconstraint respiratory measurements with load cells under the bed in awake healthy volunteers
T2 - Breath-by-breath comparison with pneumotachography
AU - Isono, Shiroh
AU - Nozaki-Taguchi, Natsuko
AU - Hasegawa, Makoto
AU - Kato, Shinichiro
AU - Todoroki, Shinsuke
AU - Masuda, Shigemi
AU - Iida, Norihito
AU - Nishimura, Toshiaki
AU - Noto, Masatoshi
AU - Sato, Yasunori
N1 - Publisher Copyright:
Copyright © 2019 the American Physiological Society.
PY - 2019/5
Y1 - 2019/5
N2 - Rate of respiration is a fundamental vital sign. Accuracy and precision of respiratory rate measurements with contact-free load cell sensors under the bed legs were assessed by breath-by-breath comparison with the pneumotachography technique during two different dynamic breathing tasks in 16 awake human adults resting on the bed. The subject voluntarily increased and decreased the respiratory rate between 4 and 16 breaths/min (n 8) and 10 and 40 breaths/min (n 8) at every 2 breaths in 6 different lying postures such as supine, left lateral, right lateral, and 30, 45, and 60° sitting postures. Reciprocal phase changes of the upper and lower load cell signals accorded with the respiratory phases indicating respiratory-related shifts of the centroid along the long axis of the bed. Bland-Altman analyses revealed 0.66 and 1.59 breaths/min standard deviation differences between the techniques (limits of agreement: 1.22 to 1.36 and 2.96 to 3.30) and 0.07 and 0.17 breaths/min fixed bias differences (accuracy) (confidence interval: 0.04 to 0.10 and 0.12 to 0.22) for the mean respiratory rates of 10.5 3.7 and 24.6 8.9 breaths/min, respectively, regardless of the body postures on the bed. Proportional underestimation by this technique was evident for respiratory rates 40 breaths/min. Sample breath increase up to 10 breaths improved the precision from 1.59 to 0.26 breaths/min. Abnormally faster and slower respirations were accurately detected. We conclude that contact-free unconstraint respiratory rate measurements with load cells under the bed legs are accurate and may serve as a new clinical and investigational tool. NEW & NOTEWORTHY Four load cells placed under the bed legs successfully captured a centroid shift during respiration in human subjects lying on a bed. Breath-by-breath comparison of the breaths covering a wide respiratory rate range by pneumotachography confirmed reliability of the contact-free unconstraint respiratory rate measurements by small standard deviations and biases regardless of body postures. Abnormally faster and slower respirations were accurately detected. This technique should be an asset as a new clinical and investigational tool.
AB - Rate of respiration is a fundamental vital sign. Accuracy and precision of respiratory rate measurements with contact-free load cell sensors under the bed legs were assessed by breath-by-breath comparison with the pneumotachography technique during two different dynamic breathing tasks in 16 awake human adults resting on the bed. The subject voluntarily increased and decreased the respiratory rate between 4 and 16 breaths/min (n 8) and 10 and 40 breaths/min (n 8) at every 2 breaths in 6 different lying postures such as supine, left lateral, right lateral, and 30, 45, and 60° sitting postures. Reciprocal phase changes of the upper and lower load cell signals accorded with the respiratory phases indicating respiratory-related shifts of the centroid along the long axis of the bed. Bland-Altman analyses revealed 0.66 and 1.59 breaths/min standard deviation differences between the techniques (limits of agreement: 1.22 to 1.36 and 2.96 to 3.30) and 0.07 and 0.17 breaths/min fixed bias differences (accuracy) (confidence interval: 0.04 to 0.10 and 0.12 to 0.22) for the mean respiratory rates of 10.5 3.7 and 24.6 8.9 breaths/min, respectively, regardless of the body postures on the bed. Proportional underestimation by this technique was evident for respiratory rates 40 breaths/min. Sample breath increase up to 10 breaths improved the precision from 1.59 to 0.26 breaths/min. Abnormally faster and slower respirations were accurately detected. We conclude that contact-free unconstraint respiratory rate measurements with load cells under the bed legs are accurate and may serve as a new clinical and investigational tool. NEW & NOTEWORTHY Four load cells placed under the bed legs successfully captured a centroid shift during respiration in human subjects lying on a bed. Breath-by-breath comparison of the breaths covering a wide respiratory rate range by pneumotachography confirmed reliability of the contact-free unconstraint respiratory rate measurements by small standard deviations and biases regardless of body postures. Abnormally faster and slower respirations were accurately detected. This technique should be an asset as a new clinical and investigational tool.
KW - Centroid
KW - Load cell
KW - Monitor
KW - Respiration
KW - Respiratory rate
UR - http://www.scopus.com/inward/record.url?scp=85065883675&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85065883675&partnerID=8YFLogxK
U2 - 10.1152/japplphysiol.00730.2018
DO - 10.1152/japplphysiol.00730.2018
M3 - Article
C2 - 30763161
AN - SCOPUS:85065883675
SN - 8750-7587
VL - 126
SP - 1432
EP - 1441
JO - Journal of Applied Physiology Respiratory Environmental and Exercise Physiology
JF - Journal of Applied Physiology Respiratory Environmental and Exercise Physiology
IS - 5
ER -