Discrimination between QRS and T Waves Using a Right Parasternal Lead for S-ICD in a Patient with a Single Ventricle

Takahiko Nishiyama, Takehiro Kimura, Nobuhiro Nishiyama, Yoshiyasu Aizawa, Keiichi Fukuda, Seiji Takatsuki

研究成果: Article査読

抄録

The subcutaneous implantable cardioverter-defibrillator (S-ICD) is a useful option for patients with a single ventricle (SV) in which transvenous leads are contraindicated because of intracardiac shunts. We report a case in which a right parasternal lead placement was indicated for an S-ICD in a resuscitated patient with an SV. There were significant changes in the magnitude of R to T waves ratio in the right compared to the left parasternal lead position. Screening in the right parasternal position is effective for selecting appropriate patients with congenital heart disease for S-ICD implantations.

本文言語English
ページ(範囲)904-907
ページ数4
ジャーナルPACE - Pacing and Clinical Electrophysiology
40
7
DOI
出版ステータスPublished - 2017 7月

ASJC Scopus subject areas

  • 循環器および心血管医学

フィンガープリント

「Discrimination between QRS and T Waves Using a Right Parasternal Lead for S-ICD in a Patient with a Single Ventricle」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。