TY - JOUR
T1 - Electrical isolation of the superior vena cava using upstream phrenic pacing to avoid phrenic nerve injury
AU - Kimura, Takehiro
AU - Takatsuki, Seiji
AU - Fukumoto, Kotaro
AU - Nishiyama, Nobuhiro
AU - Sato, Yuriko
AU - Aizawa, Yoshiyasu
AU - Fukuda, Yukiko
AU - Sato, Toshiaki
AU - Miyoshi, Shunichiro
AU - Fukuda, Keiichi
PY - 2012/9
Y1 - 2012/9
N2 - Background: Phrenic nerve (PN) injury is a potential complication that can occur during superior vena cava (SVC) isolation to cure atrial fibrillation (AF). Avoiding radiofrequency (RF) energy delivery is the safer alternative but may result in failed isolation. High-output PN pacing above the ablation site (upstream PN pacing) to confirm whether the PN is intact is a promising technique to avoid PN injury. This study was conducted to elucidate the safety of delivering RF energy at the site of capture of the right PN using upstream high-output pacing during electrical SVC isolation. Methods: SVC isolation was conducted in 41 drug-resistant AF patients. When high-output pacing (25 mA) from the distal tip of the ablation catheter captured the PN at the right atrial-SVC junction, upstream PN pacing (cycle length: 1000-1500 ms) was applied during RF delivery. The application of RF energy was stopped upon the failure or weakness of diaphragmatic twitching. The feasibility of SVC isolation using upstream PN pacing was investigated. Results: In all 41 patients, SVC isolation was successfully achieved. RF energy was delivered at the PN capture site in 26 patients (154 138 second, 18 5 W), and upstream PN pacing was successfully applied in all of the patients. Out of 46 SVC isolations, including five repeated sessions, PN injury occurred in one patient, who recovered spontaneously within 2 weeks. Conclusions: Upstream PN pacing may be effective for the safe completion of SVC isolation and to reduce the severity of PN injury. (PACE 2012; 35:1053-1060)
AB - Background: Phrenic nerve (PN) injury is a potential complication that can occur during superior vena cava (SVC) isolation to cure atrial fibrillation (AF). Avoiding radiofrequency (RF) energy delivery is the safer alternative but may result in failed isolation. High-output PN pacing above the ablation site (upstream PN pacing) to confirm whether the PN is intact is a promising technique to avoid PN injury. This study was conducted to elucidate the safety of delivering RF energy at the site of capture of the right PN using upstream high-output pacing during electrical SVC isolation. Methods: SVC isolation was conducted in 41 drug-resistant AF patients. When high-output pacing (25 mA) from the distal tip of the ablation catheter captured the PN at the right atrial-SVC junction, upstream PN pacing (cycle length: 1000-1500 ms) was applied during RF delivery. The application of RF energy was stopped upon the failure or weakness of diaphragmatic twitching. The feasibility of SVC isolation using upstream PN pacing was investigated. Results: In all 41 patients, SVC isolation was successfully achieved. RF energy was delivered at the PN capture site in 26 patients (154 138 second, 18 5 W), and upstream PN pacing was successfully applied in all of the patients. Out of 46 SVC isolations, including five repeated sessions, PN injury occurred in one patient, who recovered spontaneously within 2 weeks. Conclusions: Upstream PN pacing may be effective for the safe completion of SVC isolation and to reduce the severity of PN injury. (PACE 2012; 35:1053-1060)
KW - SVC isolation
KW - atrial fibrillation
KW - catheter ablation
KW - phrenic nerve injury
KW - phrenic pacing
UR - http://www.scopus.com/inward/record.url?scp=84866175478&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84866175478&partnerID=8YFLogxK
U2 - 10.1111/j.1540-8159.2012.03465.x
DO - 10.1111/j.1540-8159.2012.03465.x
M3 - Article
C2 - 22845419
AN - SCOPUS:84866175478
SN - 0147-8389
VL - 35
SP - 1053
EP - 1060
JO - PACE - Pacing and Clinical Electrophysiology
JF - PACE - Pacing and Clinical Electrophysiology
IS - 9
ER -