False tacrolimus concentrations measured by antibody-conjugated magnetic immunoassay in liver transplant patient: 2 case reports and literature review

Kazuaki Taguchi, Takafumi Ohmura, Yuki Ohya, Momoko Horio, Kumiko Furukawa, Hirofumi Jono, Yukihiro Inomata, Hideyuki Saito

研究成果: Article査読

5 被引用数 (Scopus)

抄録

Safe use of tacrolimus relies on regular whole-blood drug monitoring. Of the methods used to assess whole-blood tacrolimus concentration, antibody-conjugated magnetic immunoassay is mostly used for therapeutic drug monitoring because it requires only a minimal sample preparation and no pretreatment procedure. However, several cases recently have been reported in which abnormally false elevated tacrolimus concentrations were measured by antibody-conjugated magnetic immunoassay (>15 ng/mL), despite the absence of clinical symptoms. We present 2 cases of falsely detected tacrolimus concentrations that did not show abnormally high values within the therapeutic range. Whole-blood tacrolimus concentrations obtained by antibody-conjugated magnetic immunoassay showed well-controlled concentrations (approximately 2-8 ng/mL), whereas those obtained by another immunoassay and in washed erythrocytes were below the assay range (< 1.2 ng/mL). Thus, antibody-conjugated magnetic immunoassay can elicit falsely positive results of tacrolimus concentrations, even though they are within the therapeutic range.

本文言語English
ページ(範囲)474-478
ページ数5
ジャーナルExperimental and Clinical Transplantation
12
5
DOI
出版ステータスPublished - 2014 10月 1
外部発表はい

ASJC Scopus subject areas

  • 移植

フィンガープリント

「False tacrolimus concentrations measured by antibody-conjugated magnetic immunoassay in liver transplant patient: 2 case reports and literature review」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル