Measurement of the surface temperature of the cornea during ArF excimer laser ablation by thermal radiometry with a 15-nanosecond time response

Miya Ishihara, Tsunenori Arai, Shunichi Sato, Yuji Morimoto, Minoru Obara, Makoto Kikuchi

研究成果: Article査読

31 被引用数 (Scopus)

抄録

Background and Objective: The purpose of this work was to develop a temperature measurement system with a nanosecond time response to monitor the transient temperature of the corneal surface during laser refractive surgery. Materials and Methods: Thermal radiation from the surface of the porcine cornea during ArF excimer laser irradiation was measured using a photovoltaic HgCdTe detector with a response bandwidth of 150 MHz. Results: Maximum thermal radiation occurred at 31 ± 4 nanoseconds, which was longer than the time response of the measurement system. The temperature derived from the detected signal reached over 100°C at a fluence of 80 mJ/cm2, which was the ablation threshold, and reached 240°C at a fluence of 180 mJ/cm2. Conclusion: The present system of temperature measurement with a time response of 15.7 nanoseconds revealed that the transient surface temperature of the cornea during ablation is much higher than that previously reported.

本文言語English
ページ(範囲)54-59
ページ数6
ジャーナルLasers in Surgery and Medicine
30
1
DOI
出版ステータスPublished - 2002 2 6

ASJC Scopus subject areas

  • Surgery
  • Dermatology

フィンガープリント 「Measurement of the surface temperature of the cornea during ArF excimer laser ablation by thermal radiometry with a 15-nanosecond time response」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル