Monte Carlo prediction of meantime photon measurement density functions

Muneo Ono, Eiji Okada

    研究成果: Conference article査読

    抄録

    The scattering of tissue causes the ambiguity of volume of tissue interrogated by near infrared spectroscopy (NIRS) instruments. The sampling volume, which is the volume of tissue interrogated by NIRS instrument with CW light, can be predicted by calculating the distribution of light which arrives the detector. Recently, time-resolved and frequency modulated measurements has been applied to near-infrared spectroscopy (NIRS) instruments. Although the volume of tissue interrogated by the measurements of mean time, phase and so forth may be different from that by CW intensity measurement, there are very few studies on the volume of tissue interrogated by time-resolved and frequency domain measurements. In this study a Monte Carlo algorithm to predict the volume of tissue interrogated by mean time measurements is proposed and is applied to homogeneous and layered models. The volume of tissue interrogated by mean time measurement is significantly different from the sampling volume.

    本文言語English
    ページ(範囲)20-27
    ページ数8
    ジャーナルProceedings of SPIE - The International Society for Optical Engineering
    3566
    出版ステータスPublished - 1999 1月 1
    イベントProceedings of the 1998 Photon Propagation in Tissues IV - Stockholm, SWE
    継続期間: 1998 9月 91998 9月 11

    ASJC Scopus subject areas

    • 電子材料、光学材料、および磁性材料
    • 凝縮系物理学
    • コンピュータ サイエンスの応用
    • 応用数学
    • 電子工学および電気工学

    フィンガープリント