Myocardium tissue ablation with high-peak-power nanosecond 1,064- and 532-nm pulsed lasers: Influence of laser-induced plasma

Makoto Ogura, Shunichi Sato, Miya Ishihara, Satoko Kawauchi, Tunenori Arai, Takemi Matsui, Akira Kurita, Makoto Kikuchi, Hiroshi Ashida, Minoru Obara

研究成果: Article

9 引用 (Scopus)

抜粋

Background and Objectives: We investigated the mechanism and characteristics of porcine myocardium tissue ablation in vitro with nanosecond 1,064- and 532-nm pulsed lasers at laser intensities up to ≈5.0 GW/cm2. Particular attention was paid to study the influence of the laser-induced plasma on the ablation characteristics. The applicability of these two lasers to transmyocardial laser revascularization (TMLR) was discussed. Study Design/Materials and Methods: Porcine myocardium tissue samples were irradiated with 1,064- and 532-nm, Q-switched Nd:YAG laser pulses, and the ablation depths were measured. The temporal profiles of the laser-induced optical emissions were measured with a biplanar phototube. For the ablated tissue samples, histological analysis was performed with an optical microscope and a polarization microscope. Results: The ablation efficiency at 1,064 nm was higher than that at 532 nm. The ablation threshold at 1,064 nm (≈0.8 GW/cm2) was lower than that at 532 nm (≈1.6 GW/ cm2), in spite of the lower absorption coefficient being expected at 1,064 nm. For the 1,064-nm laser-ablated tissues, thermal damage was very limited, while damage presumably caused by the mechanical effect was observed in most of the cases. For the 1,064-nm laser ablation, the ablation threshold was equal to the threshold of the laser-induced optical emission (≈0.8 GW/cm2), while for the 532-nm laser ablation, the optical emission threshold (≈2.4 GW/cm2) was higher than the ablation threshold. Conclusions: We considered that for the 1,064-nm laser ablation, the tissue removal was achieved through a photodisruption process at laser intensities of > ≈0.8 GW/cm2. At laser intensities of > 3.0 GW/cm2, however, the ablation efficiency decreased; this can be attributed to the absorption of incoming laser pulses by the plasma. For the 532-nm laser ablation, the tissue removal was achieved through a photothermal process at laser intensities of > ≈1.6 GW/ cm2. At laser intensities of > 2.4 GW/cm2, a photodisruption process may also contribute to the tissue removal, in addition to a photothermal process. With regard to the ablation rates, the 1,064-nm laser was more suitable for TMLR than the 532-nm laser. We concluded that the 1,064-nm Q-switched Nd:YAG laser would be a potential candidate for a laser source for TMLR because of possible fiber-based beam delivery, its compact structure, cost effectiveness, and easy maintenance. Animal trials, however, have to be carried out to evaluate the influence of the tissue damage.

元の言語English
ページ(範囲)136-141
ページ数6
ジャーナルLasers in Surgery and Medicine
31
発行部数2
DOI
出版物ステータスPublished - 2002 9 17

ASJC Scopus subject areas

  • Surgery
  • Dermatology

フィンガープリント Myocardium tissue ablation with high-peak-power nanosecond 1,064- and 532-nm pulsed lasers: Influence of laser-induced plasma' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用

    Ogura, M., Sato, S., Ishihara, M., Kawauchi, S., Arai, T., Matsui, T., Kurita, A., Kikuchi, M., Ashida, H., & Obara, M. (2002). Myocardium tissue ablation with high-peak-power nanosecond 1,064- and 532-nm pulsed lasers: Influence of laser-induced plasma. Lasers in Surgery and Medicine, 31(2), 136-141. https://doi.org/10.1002/lsm.10081