New Noninvasive Transcutaneous Approach to Blood Glucose Monitoring: Successful Glucose Monitoring on Human 75 g OGTT with Novel Sampling Chamber

Shinsuke Kayashima, Tsunenori Arai, Makoto Kikuchi, Noriharu Sato, Naokazu Nagata, Osamu Takatani, Narushi Ito, Jun Kimura, Toshihide Kuriyama, Akio Kaneyoshi

研究成果: Article査読

29 被引用数 (Scopus)

抄録

A novel noninvasive and quasi-continuous method of transcutaneous blood glucose monitoring for use with the human 75 g oral glucose tolerance test (OGTT) has been developed. The effused fluid was obtained by applying suction on the skin surface and labeled suction effusion fluid (SEF). The system consists of two main parts: a suction apparatus and the glucose sensor system. The suction apparatus applies vacuum to the patient’s skin at 400 mmHg absolute pressure to collect the SEF. The miniature ion sensitive field effect transistor (ISFET) based glucose sensor can measure glucose in small SEF quantities. The monitoring system is based on the association between the glucose concentration in the SEF and in the serum. During the 75 g OGTT, the glucose change in the SEF was measured every 10 min. Although a response delay of up to 20 min was observed in the SEF glucose change, it was possible to perform the 75 g OGTT by this noninvasive monitoring method.

本文言語English
ページ(範囲)752-757
ページ数6
ジャーナルIEEE Transactions on Biomedical Engineering
38
8
DOI
出版ステータスPublished - 1991 8
外部発表はい