Ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers with segmental or equivalent refractive indices

Miki Kamikawatoko Omoto, Hidemasa Torii, Sachiko Masui, Masahiko Ayaki, Kazuo Tsubota, Kazuno Negishi

研究成果: Article

4 引用 (Scopus)

抜粋

This study compared the axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), mean anterior corneal radius of curvature (Rm), and postoperative refractive outcomes obtained from two different swept-source optical coherence biometers, the ARGOS (Movu, Nagoya, Japan), which uses the segmental refractive index for each segment, and the IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany), which uses an equivalent refractive index for the entire eye. One hundred and six eyes of 106 patients with cataracts were included. The refractive outcomes using the Barrett Universal II, Haigis, Hoffer Q, and SRK/T formulas were evaluated. The mean AL, CCT, ACD, and Rm differed significantly (P < 0.001) with the IOLMaster 700 (25.22 mm, 559 µm, 3.23 mm, and 7.69 mm) compared with the ARGOS (25.14 mm, 533 µm, 3.33 mm, and 7.66 mm). The mean LTs did not differ significantly. The percentages of eyes within ±0.50 and ±1.00 diopter of the predicted refraction did not differ significantly (P > 0.05). The accuracy of the intraocular lens power calculations was clinically acceptable with both biometers, although the ocular biometry using these two biometers exhibited certain differences.

元の言語English
記事番号6557
ジャーナルScientific reports
9
発行部数1
DOI
出版物ステータスPublished - 2019 12 1

ASJC Scopus subject areas

  • General

フィンガープリント Ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers with segmental or equivalent refractive indices' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用