抄録
CE-TOFMS is a powerful method for profiling charged metabolites. However, the limited availability of metabolite standards hinders the process of identifying compounds from detected features in CE-TOFMS data sets. To overcome this problem, we developed a method to identify unknown peaks based on the predicted migration time (tm) and accurate m/z values. We developed a predictive model using 375 standard cationic metabolites and support vector regression. The model yielded good correlations between the predicted and measured tm (R=0.952 and 0.905 using complete and cross-validation data sets, respectively). Using the trained model, we subsequently predicted the tm for 2938 metabolites available from the public databases and assigned tentative identities to noise-filtered features in human urine samples. While 38.9% of the peaks were assigned metabolite names by matching with the standard library alone, the proportion increased to 52.2%. The proposed methodology increases the value of metabolomic data sets obtained from CE-TOFMS profiling.
本文言語 | English |
---|---|
ページ(範囲) | 2311-2318 |
ページ数 | 8 |
ジャーナル | ELECTROPHORESIS |
巻 | 31 |
号 | 14 |
DOI | |
出版ステータス | Published - 2010 7月 |
ASJC Scopus subject areas
- 分析化学
- 生化学
- 臨床生化学