Removing ambiguity caused by T2 shine-through using weighted diffusion subtraction (WDS)

Koichi Oshio, Shigeo Okuda, Hiroshi Shinmoto

研究成果: Article

1 引用 (Scopus)

抜粋

We propose a novel image processing technique that combines images routinely acquired with low and high b values to create a single image that contains clinically useful information without the ambiguity of T2 shine-through. The contrast of resulting images is similar to that of a T2 image, but the signals of pixels with low apparent diffusion coefficient (ADC) values are inverted. The proposed technique takes the threshold ADC value as the one adjustable parameter.

元の言語English
ページ(範囲)146-148
ページ数3
ジャーナルMagnetic Resonance in Medical Sciences
15
発行部数1
DOI
出版物ステータスPublished - 2016

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

これを引用