Safety of argon plasma coagulation for hemostasis during endoscopic mucosal resection

Mitsuhiro Fujishiro, Naohisa Yahagi, Masanori Nakamura, Naomi Kakushima, Shinya Kodashima, Satoshi Ono, Katsuya Kobayashi, Takuhei Hashimoto, Nobutake Yamamichi, Ayako Tateishi, Yasuhito Shimizu, Masashi Oka, Masao Ichinose, Masao Omata

研究成果: Article査読

13 被引用数 (Scopus)

抄録

Showing the safety of argon plasma coagulation (APC) over mucosal defects during/after endoscopic mucosal resection (EMR), 2 studies using resected pig (ex vivo) and living minipig (in vivo) stomachs were performed. As an ex vivo study, APC was applied over mucosal defects in 2 groups; with prior submucosal saline injection and without injection. Only subtle tissue damage was observed in the injection group, whereas apparent damage was observed in the noninjection group. The damaged distances in depth significantly increased as the pulse duration increased and those at the pulse duration of 4 seconds, which might be maximal in clinical practice, were approximately 1 mm. As an in vivo study, APC was applied over mucosal defects immediately after EMR. Only subtle tissue damage was observed even at the pulse duration of 20 seconds as shown in the ex vivo study. APC can be performed safely over the mucosal defects during/after EMR.

本文言語English
ページ(範囲)137-140
ページ数4
ジャーナルSurgical Laparoscopy, Endoscopy and Percutaneous Techniques
16
3
DOI
出版ステータスPublished - 2006 6
外部発表はい

ASJC Scopus subject areas

  • 外科

フィンガープリント

「Safety of argon plasma coagulation for hemostasis during endoscopic mucosal resection」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル