抄録
The zero detectin threshold of a liquid scintillation spectrometer is found to be 2.4 ± 0.10 keV below zero pulse-height by measuring standardized 3H samples with a commercially-available liquid-scintillation spectrometer. Based on the results, a high accuracy modified integral counting method which extrapolates the integral pulse-height spectrum to the zero detection threshold is proposed.
本文言語 | English |
---|---|
ページ(範囲) | 341-344 |
ページ数 | 4 |
ジャーナル | Applied Radiation and Isotopes |
巻 | 45 |
号 | 3 |
DOI | |
出版ステータス | Published - 1994 3月 |